Safety notice 
1. Transportation 
As microscope is a precision instrument, handle with care,   
avoiding impact or abrupt movement during transportation.   
Do not push or pull the microscope during using, otherwise   
the precision for image will be reduced. 
1.1 Hold the curve and keep the microscope in balance 
1.2 Do not hold the focusing knob, eyepiece tubes and stage   
as these parts are movable. Troubles maybe caused by such   
handles. 
Do not make specimen or filters fall off. 
 
2. Working Environment 
As microscope is a precision instrument, improper using will make it unworkable or reduce its precision.   
2.1 Do not expose the microscope in the sun directly 
2.2 Temperature range is 0

℃~

40

 and the max. humidity is 85% 

2.3 Avoid high temperature and humidification otherwise there will be fog or mold on the lens 
2.4 Avoid violent vibration as the vibration will reduce the image’s quality 
2.5 Place microscope on a stable plane and keep it in balance. Keep the environment breezy and cover the 
microscope with the dust-cover 
2.6 Do not place microscope in moist room to avoid short circuit. Please turn off the power supply as soon 
as water drop in the microscope. If there are other things unsafe come into the microscope may cause short 
circuit too, please stop using and contact with manufacturers 
 
3. Focusing knob 
Never turn the left and right focusing knob in the adverse   
direction at the same time. Do not turn the coarse focusing   
knob when the stage reaches max. position. These improper   
operations will make focusing structure unworkable. 
 
 
4. Clamp 
The clamps fix the microscope firmly during transportation.   
Please remove them before using. 
 
 
 
 
5. Do not take the microscope apart as it will be damaged. Such operations might have bad effect on the 
performance and make user get an electric shock or injured. Please contact with manufacturers if there is 
any problem. 
 
6. There is correct input voltage in the label on microscope.   
Make sure the voltage in your position is right. Improper input   

clamp 

Содержание A12.1010

Страница 1: ...A12 1010 Laboratory Biological Microscope Instruction Manual Please read this manual carefully before use microscope...

Страница 2: ...Contents BM2000 1 Structure and nomenclature 4 2 Technical parameter and specification 6 3 Configuration 8 4 Assemblage 9 5 Operation 13 6 Troubleshooting 18 7 Maintenance 20 8 Products Standards 20...

Страница 3: ...oscope on a stable plane and keep it in balance Keep the environment breezy and cover the microscope with the dust cover 2 6 Do not place microscope in moist room to avoid short circuit Please turn of...

Страница 4: ...inserted into the holder firmly Such Operation may cause short circuit or smoking At last please make sure if the collector has been mounted properly 9 Temperature for illumination There will be high...

Страница 5: ...re screw Nosepiece Objectives Specimen holder Mechanical stage Condenser with auxiliary lens Collector Binocular head Eye tubes fixture Microscope body Power supply Lightness adjusting knob switch scr...

Страница 6: ...Working environment Temperature 0 40 Humidity max 85 no dew Altitude within 2000m Pollution 2 Assembly 2 Optical parameter 2 1 Objectives infinity achromatic objectives Magnification N A Cover glass t...

Страница 7: ...tem Content Piece 1 Microscope s body including stage nosepiece coaxial coarse and fine focusing adjustment bracket for condenser adjusting power supply halogen bulb and normal collector 1pc 2 Siedent...

Страница 8: ...ded AC connector and make sure that the electronic cord has been connected safely Please use the supplied electronic cord by manufacturer Please make sure to use PE electronic cord when using extra lo...

Страница 9: ...rsion image disappeared 4 5 2 Objectives The objectives have been mounted in the microscope before shipment The objectives have been mounted in the microscope before shipment Take off the specimen fro...

Страница 10: ...lacement for fuse 4 6 2 1 Turn off the power supply turn the switch to O and pull out the connector 4 6 2 2 Open the cover for fuse by screwdriver 4 6 2 3 Mount the new fuse 4 6 2 4 Make sure that the...

Страница 11: ...there is sharp image 5 4 4 Turn the fine focusing knob to focus finely Use the 10X objective or 4X objective firstly before 40X and 100X objective Turn the fine focusing knob to focus finely 5 5 Adjus...

Страница 12: ...ctive when the aperture diaphragm adjusting shaft reaches some position If the objective has been changed please move the shaft to the position with the same value which is in the objective in the opt...

Страница 13: ...4 2 Never press the bottle overly as it may caused oil gushed out of the bottle 5 9 4 3 Clean the out surface of the bottle from remanent oil 5 10 Adjusting the coarse focusing knob s tension The tens...

Страница 14: ...or dust on the slide Cleaning Condenser position too low Correct position Poor image quality low resolution poor contrast No cover glass on the specimen Attach cover glass Cover glass too thick or thi...

Страница 15: ...Inadequate brightness or illumination Adjust brightness with control dial 3 Electrical Trouble Cause Remedy Lamp dose not light when switched ON No electrical power Check power cord connection Lamp b...

Страница 16: ...er thinner ect to clean the painted or plastic surfaces of the instrument We recommend the use of silicon cloth More persistent dirt may be cleaned with mild detergent solution Printed plastic surface...

Отзывы: