Keratron
™
Piccolo Installation and operating manual
OPTIKON 2000
9 TROUBLESHOOTING
GUIDE
9.1 SOFTWARE INSTALLATION
After the installation check the proper working of software running and test Scout
program. If any error messages appears (Automation error, DLL xxx.dll not fount etc)
the suggested actions are:
•
Unload the Scout software (see §6.4.2)
•
Reinstall it overwriting the files when prompted.
9.2 THE SCOUT.INI FILE
This files contain all the settings you make using the “Option Panel”, and other
settings provided by the SW.
This file must be used if::
1. The SW hangs when starts.
When this occurs try to rename (
not delete!!
) the “Scout.ini” file and restart the
SW..
2. Runtime error exporting tests.
When this occurs open the Scout.ini file and then, in the “[Export]” section, check
the “path” key, probably it is pointing to a network path not available at the
moment, or to an external drive not connected. Delete this row
“path=x:\nnnn\mmmm\” and retry.
9.3 SHARING KERATRON SCOUT WITH OTHERS
If more than one user performs tests with Keratron
™
Piccolo, how can I separate
tests to avoid confusion ?
There are two ways to proceed:
•
Create your own “group” (e.g. your name) into the database, and save every
test you perform in that “group”. (
not recommended
)
Cod. 161301EN
9-1
2012-01-17
Rev.D