OPTIKON I.V. Скачать руководство пользователя страница 19

I.V. Pole cart 

Installation and operating manual 

OPTIKON 2000 

Code 181004EN 

 

2007-10-09 Rev. 0 

 

6-3

d.  Pass the connector of the control panel's main footswitch through the large 

hole (from inside) located on the lower right side of the cart (Fig. 2). Then 
connect the I.V. pole's Up/Down footswitch to the "footswitch" connector (Fig. 
1, no. 3) by passing it through the large hole (from inside) located on the 
lower right side of the cart (Fig. 2). 

e.  Connect the power cable provided with the cart 181004  to the appropriate 

mains inlet (Fig. 1, no. 4) and lock it in place using the strain-relief safety 
system, make sure that the voltage setter (Fig. 1, no. 5) is set on the proper 
level of voltage for the mains power, and then remount the protection cover. 
Lastly, connect the power cable to the power outlet in the operating room. 

f.  Switch on the equipment and the cart, then proceed according to the hosted 

equipment user manual. 

6.4 SHUTDOWN 

PROCEDURE 

If the Optikon cart is not being used for an extended period, the following procedure should 
be observed:  
 

1. 

Set the pole to the minimum height. 

2. 

Depress the “Power” switch on the rear of the cart column to power 

down the console and the cart. 

3. 

Disconnect the cart's cable from the power outlet and wrap it around 

the column. Be sure to pull the plug not the cord. 

4. 

Place the cart footswitch and the system footswitch inside the cart's 

main drawer. 

5. 

Cover the hosted equipment and the cart with a drape to protect from 

dust. 

 

6.5 MAINTENANCE 

The 181004 Optikon Equipment Cart requires very few attentions to provide you with a 
long lasting satisfactory service. 
The medical personnel must take the responsibility to keep existing instruments and 
equipment functioning properly. The following simple steps serve as practical guidelines in 
establishing your care and maintenance program: 

1.  The Optikon Equipment Cart and the hosted console(s) should be stored in a clean 

and dry environment at room temperature. 

2.  In order to remove salt solution deposits or other residuals, periodically clean the 

cart and the control console with soft dusting cloth dampened with distilled water. If 
necessary use only neutral detergent. 

3.  Do not move the cart or equipment footswitch by pulling on their respective cables. 
4.  Do not overload the mayo tray and/or the equipment tray. 

 
 
 

Содержание I.V.

Страница 1: ...Rev 0 EQUIPMENT CART WITH ELECTRIC I V POLE INSTALLATION AND OPERATING MANUAL OPTIKON 2000 S p A Via del Casale di Settebagni 13 00138 Rome Italy Tel 39 06 88884355 Fax 39 06 8888440 e mail sales opt...

Страница 2: ...O 9001 and ISO 13485 certified company that manufactures surgical and diagnostic devices for ophthalmology All OPTIKON 2000 products are manufactured in compliance with the requirements of Directive 9...

Страница 3: ...ATION 5 1 5 1 SYSTEM DESCRIPTION 5 1 5 2 CONTROLS AND INDICATORS 5 1 5 3 TECHNICAL SPECIFICATIONS 5 3 5 4 CIRCUIT DIAGRAMS 5 4 6 INSTALLATION AND OPERATING PROCEDURE 6 1 6 1 INTRODUCTION 6 1 6 2 INITI...

Страница 4: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...

Страница 5: ...f updates calibrations and repairs are carried out by personnel which has been authorised by OPTIKON 2000 S p A the system is used in compliance with the user instructions the mains electric installat...

Страница 6: ...I V Pole cart Installation and operating manual OPTIKON 2000 Code 181004EN 2007 10 09 Rev 0 1 2 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...

Страница 7: ...made by persons who are not part of the authorised OPTIKON S p A technical service OPTIKON 2000 S p A reserves the right to verify in case of failures if the instrument and or its accessories have be...

Страница 8: ...I V Pole cart Installation and operating manual OPTIKON 2000 Code 181004EN 2007 10 09 Rev 0 2 2 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...

Страница 9: ...OF PROFESSIONAL MEDICAL JUDGMENT IN ALL CASES OPTIKON 2000 S p A IS NOT RESPONSIBLE FOR ANY CLINICAL PROBLEM RESULTING FROM MISHANDLING THIS INSTRUMENT AND MAKES NO MEDICAL RECOMMENDATIONS THE SYSTEM...

Страница 10: ...UM OF APPROX 235cm OF HEIGHT DO NOT PLACE THE CART UNDER LOW CEILING DO NOT TILT THE CART MORE THAN 10 IN ANY DIRECTION TO AVOID RISK OF SMOKE OR FIRE WHEN PLUGGING HOSTED EQUIPMENT TO THE CART DO NOT...

Страница 11: ...e The user should become acquainted to these symbols before using the equipment SYMBOL DESCRIPTION ALTERNATING CURRENT DIRECT CURRENT PROTECTIVE EARTH ATTENTION CONSULT ACCOMPANYING DOCUMENTS OFF POWE...

Страница 12: ...I V Pole cart Installation and operating manual OPTIKON 2000 Code 181004EN 2007 10 09 Rev 0 4 2 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...

Страница 13: ...dure An electric I V pole holds the BSS bottle to be used as irrigation source The bottle position height can be adjusted by the means of a rocker footswitch and or it can be controlled by the connect...

Страница 14: ...een installed The following are located under the protection cover Fig 1 1 MAINS POWER CONNECTION FOR HOSTED CONSOLES Three mains sockets allow connecting the mains power for the hosted surgical conso...

Страница 15: ...1 1 Classification Class I ELECTRICAL SPECIFICATIONS Input voltage 220 V 50 Hz 110 V 50 60 Hz Fuses 220 V 6 3 A T 110 V 16 A T IV Pole power consumption 40 VA Maximum connectable power 1 KVA I V POLE...

Страница 16: ...2 Kg IMPORTANT NOTICE The footswitch is not autoclaveable The switches located inside the footswitch are immersion proof protection class IPX8 Therefore this accessory is suitable for use in operating...

Страница 17: ...g instructions carefully in a step by step manner and within a short time you will have gained all the knowledge needed for successful operation of your surgery system 6 2 INITIAL INSPECTION This inst...

Страница 18: ...n b Place the Optikon Console on the equipment tray c Remove the protection cover from the electric sockets by unscrewing the 4 socket head screws Removing the protection cover will give you access to...

Страница 19: ...ole to the minimum height 2 Depress the Power switch on the rear of the cart column to power down the console and the cart 3 Disconnect the cart s cable from the power outlet and wrap it around the co...

Страница 20: ...OTE OPTIKON 2000 S p A reserves the right to change design and specifications without notice because of its policy of continuous product research and development CAUTION DO NOT USE SYNTHETIC DETERGENT...

Страница 21: ...OMS CORRECTIVE ACTION Mains power supply System completely inoperative a A C power cord not connected Connect power cord to the cart and the power source b A C line fuse blown Replace the line fuses c...

Отзывы: