I.V. Pole cart
Installation and operating manual
OPTIKON 2000
Code 181004EN
2007-10-09 Rev. 0
5-2
POWER (ON/OFF MAINS SWITCH/PILOT LAMP)
This illuminated switch is located on the left side of the cart and controls both the internal
electric I.V. pole and the power outlets for the control panels on the cart. Provided that the
mains switches of the single consoles are left to the on position, this switch will energize or
de-energize all the surgical system.
PROTECTION COVER
This protection cover, located at the rear of the cart, was designed to prevent the use of
the cart's power outlets to connect any device other than approved Optikon instruments.
The lid must be removed by a qualified technician when the
console is first installed
.
The
lid must be restored in place once the equipment has been installed
. The following are
located under the protection cover (Fig. 1):
1.
MAINS POWER CONNECTION FOR HOSTED CONSOLES.
Three mains sockets allow connecting the mains power for the hosted surgical
consoles.
2.
INTERFACE CONNECTOR
It allows the position control of the integral electric I.V. pole by the hosted
console. The console connects here by the means of the signal cable supplied
with the cart.
3.
FOOTSWITCH CONNECTOR
The bottle Up/Down rocker footswitch connects here.
4.
MAINS INLET AND FUSE HOLDER
The power cord to energize the cart must be connected to this inlet. Use only the
supplied cord or a Medical grade power cord. The socket also hosts two fuses that
interrupt the mains power in case of fault in the cart or in the hosted consoles.
5.
VOLTAGE SETTING CONNECTOR
By turning the voltage setting connector to the proper position, it is possible to set
the mains voltage (220 or 110 V).
TAG (MARKINGS
)
Provides serial number, electrical rating and other information as required by EN 60601-1