The instrument is electrically powered. Please ensure that the correct voltage
settings have been applied before applying power to the instrument. If in doubt
consult a qualified electrician. The instrument has a rating label affixed to the rear.
Please consult this if needed.
Always disconnect the equipment before moving or removing any guards or
covers. Switch off at the mains, remove the mains plug from the wall socket and
remove the cable from the inlet socket on the rear.
While every effort has been made to protect the inside of the instrument
against splashes, the instrument carries no IP rating. If fluids are spilt on the
instrument they may cause damage and cause an electrical hazard.
If a spill occurs, remove power from the instrument. Do not touch the
instrument or any fluid flowing from it while it is connected to the mains supply.
Always follow local health and safety guidelines.
The instrument is not intended for use with flammable substances.
It is not advised to open the lids whilst the instrument is running as it could
cause measurement errors leading to incorrect results.
It is advised to run the ‘Self-Test’ function at the start of every session. This is
located in Settings.
OptiGene Limited recommends that every two years the instrument's
performance is checked for degradation. This can be done by requesting and
purchasing ready-made strips from OptiGene that will perform an amplification
and anneal run on the instrument which can then be assessed to check for
optical or thermal performance issues.
Normal safe local operating standards should be applied at all times. The warnings above are
for guidance only. Please consult the instrument supplier if there is any doubt.