IFU_EN_GEN-HT Version 1.01
July2020
42
SHOW PLOT
Selecting ‘Show plot’ will produce a second
menu to select which plot to display on the
screen depending on which signal
processing step is wanting to be viewed.
These are explained below and further in
Chapter 6.
Raw
Raw unprocessed fluorescence
Spike
The data after spike removal
Step
The data after step removal
Slope
The data after slope correction
Normalised
The data after normalisation
Log
The log of the normalised data.
Ratio
The ratio (dF/F) of adjacent points (after step removal). This plot is
smoothed with an averaging filter.
Derivative
The gradient of the data (generated with a differentiating filter)
2nd Derivative
The gradient of the derivative (a second application of the same
filter)
Ratio4
An alternative ratio ((F-1-F1)/F0²) that gives an earlier indication of
amplification.
RESULT CALLING
Result calling is explained in further detail
in Chapter 6.