b.
Automated ISO Setup
method
i.
No additional inputs are required when using this
method. The software performs an initial scan of the
beam and populates either the
Step Table
or
Automated Stepping
values with the points
recommended to achieve an ISO measurement.
Automated ISO Setup
is not available in Manual mode.
c.
Four Cuts
method
Choosing Four Cuts Values
i.
The
Z1
should be a point selected somewhere
before the waist.
ii.
The
Z2
should be a point beyond the waist where
the beam width is smaller than at
Z1
.
iii.
The
Z3
should be a point selected beyond the waist,
and should be a value close to
Z2
where the beam
width of
Z1
falls between the widths at
Z2
and
Z3
.
iv.
The remaining cuts are populated automatically by
the software.
If these criteria are not met, the wizard fails and an error
appears, terminating the run. Should this occur, reenter
proper values.
During a run, the software usually prompts the user to collect point
Z6
after
Z1
. The run fails if
Z2
is taken instead of
Z6
.
The Four Cuts method is not ISO standard, but can supply quicker results that are just as accurate.