Run Ribbon
3.3
The
Run
ribbon
contains controls for
operation.
Live Playback
3.3.1
When enabled, the
2D Beam Display
is shown in real-time. This feature provides a laser
alignment aid to verify the setup of the optical system.
Data frames collected in live mode are not stored and therefore cannot be saved or loaded.
Results in the
Frame (Quantitative)
group are computed, but an Ultracal cycle cannot be
performed during Live Playback, so results may not be accurate. Nevertheless, the Peak Energy
level is useful in setting external attenuation such that the dynamic range of the camera is utilized.
Start
3.3.2
Use
Start
to begin an M
2
run with the last programmed or loaded settings. View or change the
settings in the
Configuration
wizard, accessed through the
Configure
button.
Configure
3.3.3
Use
Configure
to open the
Configuration
wizard, input information about the system under
test, select the operating mode (Automated Stepping, Automated ISO Setup, Four Cuts, or Step
Table), and setup the measurement parameters. See section 4.2.1 for information on operation.
Info
3.3.4
Use
Info
to open only the
Run Info
window of the
Configuration
wizard. View and edit the model number, serial number, and
comments for the next run. Changes in this window and the
Configuration
wizard are shared.