background image

 

HOBO RX3000 Remote Monitoring Station Manual 

Specifications (continued) 

 

The CE Marking identifies this product as complying with all relevant 
directives in the European Union (EU) 

 

 

 

 

See last page 

RX3002:

 FCC ID R68XPICOW, IC ID 3867A-XPICOW 

RX3003:

 FCC ID QIPEHS6, IC ID 7830A-EHS6; approved for use in 

Taiwan and Japan 

RX3004:

 FCC ID QIPPLS62-W, IC ID:7830A-PLS62W 

Ethernet (RX3001) 

 

Connector 

One RJ45/100BaseT 

Wi-Fi (RX3002) 

 

Network Standards 

IEEE 802.11b/g/n 

Frequency Range 

2.412–2.484 GHz 

Antenna Connector 

1, no diversity supported 

Data Rates 

1, 2, 5.5, 11 Mbps (802.11b); 6, 9, 12, 18, 24, 36, 48, 54 Mbps (802.11g
802.11n, HT20 MCS0 (6.5 Mbps) to HT20 MC87 (65 Mbps) 

Number of Selectable Radio 
Subchannels 

Up to 14 channels; profiles available will include USA, France, Japan, 
Spain, Canada, and “Other” (multiple countries) 

Radio Modulations 

OFDM, DSSS, DBPSK, DQPSK, CCK, 16QAM, 64QAM 

Security 

WEP 64/128, WPA-PSK, AES end-to-end encryption 

Maximum Receive Level 

-10 dBm (with PER <8%) 

Receiver Sensitivity 

-72 dBM for 54 Mbps, -87 dBm for 11 Mbps, -89 dBm for 5.5 Mbps,  
-90 dBm for 2.0 Mbps, -92 dBm for 1.0 Mbps 

Cellular (RX3003 and RX3004) 

 

Wireless Radio 

RX3003: 

GSM/GPRS/EDGE: Quad band 850/900/1800/1900 MHz, 
UMTS/HSPA+: Five band 800/850/900/1900/2100 MHz 

RX3004:  

GSM/GPRS/EDGE: Quad band 850/900/1800/1900 MHz 
UMTS/HSPA+: Seven band 800/850/900/1800/1900/2100 MHz 
LTE: Twelve Band 700/800/850/900/1800/1900/2100/2600 MHz 

Antenna 

RX3003:

 Penta band 

RX3004:

 4G LTE 

Optional Analog Sensor Module (RXMOD-A1)

 

Input Channels 

Four, single-ended, in addition to smart sensor data channels 

Measurement Range and 
Accuracy 

0–25.6 mA DC, ±5 µA ± 0.15% of reading 
0–2.5 V DC, ±0.25 mV ±0.2% of reading 
0–5 V DC, ±0.25 mV ±0.2% of reading 
0–10 V DC, ±0.3 mV ±0.2% of reading 
0–20 V DC, ±0.6 mV ±0.2% of reading 
0–33 V DC, ±1.20 mV ±0.2% of reading 

Resolution 

15 bits 

Field Wiring 

Two- or three-wire via screw terminals, 16–24 AWG 

Minimum/Maximum Input 
Voltage 

0/33 V DC 

Minimum/Maximum Input 
Current 

0/25.6 mA 

Minimum Source Impedance 
for Current Measurement 

20 KΩ 

Excitation Voltage 

12 V DC ±5% at 200 mA maximum per module 

Alarm Output Relays 

Each relay contact closure can be configured as normally open, 
normally closed, or pulsed for one second 

Voltage 

30 V max 

 

 

 

 

Dimensions 

 

 

Analog Sensor Module (RXMOD-A1) 

11.8 cm 

4.7 in. 

18.6 cm 
(7.3 in.) 

18.1 cm 

(7.1 in.) 

www.

.com

[email protected]

1.800.561.8187

Содержание HOBO RX3000 Series

Страница 1: ...ns AC power adapter AC U30 solar panel SOLAR xW or external power source 5 V DC to 17 V DC with external DC power cable CABLE RX PWR Rechargeable Battery Service Life Typical 3 5 years when operated i...

Страница 2: ...dBm for 1 0 Mbps Cellular RX3003 and RX3004 Wireless Radio RX3003 GSM GPRS EDGE Quad band 850 900 1800 1900 MHz UMTS HSPA Five band 800 850 900 1900 2100 MHz RX3004 GSM GPRS EDGE Quad band 850 900 180...

Страница 3: ...s product as complying with all relevant directives in the European Union EU RXMOD RXW 922 See last page Table of Contents Device Components and Operation 4 LCD Operation 5 Setting up the Station 7 a...

Страница 4: ...ation Charging Port Use this port to plug in an AC adapter solar panel or external power source to keep the battery charged see Setting up the Station and Battery Information USB Port Use this port to...

Страница 5: ...dicates the last connection to HOBOlink was ok indicates there was a problem with the last connection check the Connections log in HOBOlink Smart Sensor and Module Status This part of the LCD shows th...

Страница 6: ...losed c In this example the first and third relays are open and the second one is closed This will blink in the lower right part of the LCD when a firmware update is underway It will display which mod...

Страница 7: ...er module on the left and the relay module on the right for easier cable routing c Using a Phillips head screwdriver tighten the screw at the bottom of each module In this example an analog module is...

Страница 8: ...eckbox f Enter the IP Address Subnet Mask Gateway and DNS Server Consult your Network Administrator for the appropriate addresses to complete these fields g Click Save in the RX3000 Manager Click Done...

Страница 9: ...or one smart sensor through the larger of the two cable openings and plug it into one of the 10 smart sensor connectors Repeat for any additional smart sensors b Press the Select button to view the sm...

Страница 10: ...ion 4 GND EX RTN 4 GND EX RTN 5 CH2 SIGNAL 5 CH4 SIGNAL 6 CH2 GND 6 CH4 GND 7 SHIELD 7 SHIELD Note All four input channels share the same common ground Analog Module Functional Diagram To connect rela...

Страница 11: ...section on your station s page in HOBOlink c Click Save or click Next Smart Sensors Logging and Configuration You can configure both the global settings that affect all smart sensors logging interval...

Страница 12: ...he sensor measurement type for that mote g Click one of the mote measurement types from menu under Module Wireless Sensors Logging as shown in the following example h Type a label for the measurement...

Страница 13: ...he four analog sensor channels from the menu such as Channel 1 shown in this example g Select Enable Graph if you want the sensor data to be graphed in HOBOlink h Select Enable this channel if you wan...

Страница 14: ...ired 4 Under Select Data to Export choose the name of the custom data export you want to be delivered or follow the previous set of steps to set up a custom data export 5 Under Data Destination select...

Страница 15: ...ps send an email or text or close open or pulse one of the three relays if a relay module is installed For email or text enter the details and then select Send on Clear Also if you want an email or te...

Страница 16: ...is inserted through the bottom of the station case Do not reconnect the power on the station until the mote is plugged in as shown 7 Plug in the battery and charging device and wait for the device to...

Страница 17: ...ocess make sure the mote is in a vertical upright position and within range of the station 7 Repeat steps 4 6 for any additional motes to add 8 Press the Search button the magnifying glass on the RX30...

Страница 18: ...t in the connection schedule Also note that the station will connect to HOBOlink when the device is powered up and when you press the Start button For RX3003 and RX3004 cellular models All connections...

Страница 19: ...location Check the signal strength on the mote LCD on the location where you want to place the mote If there is only one or two bars on the signal strength indicator consider moving the mote to a loca...

Страница 20: ...d for outdoor and weatherproof deployments and recommended for harsh indoor environments where debris could enter the station 1 Make sure all sensors and cables are installed including the solar panel...

Страница 21: ...n or mote before opening it Check that all cables and wires are free of damage such as cracks cuts and splits Make sure cables and wires are still fastened securely and any conduit is still intact Ver...

Страница 22: ...ranteed due to uncertainties in initial battery conditions and operating environment Use the following table as a guideline for estimating how long the station will run using a fully charged battery a...

Страница 23: ...e separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV t...

Страница 24: ...minutes in air moving 1 m sec with protective cap Operating Temperature Range 40 C to 75 C 40 F to 167 F Environmental Rating Weatherproof 0 to 100 RH intermittent condensing environments For best re...

Страница 25: ...mounted properly the housing is weatherproof but not waterproof Refer to the station manual and Tripod Setup Guide for more information regarding setting up stations Connecting the Sensor to a Statio...

Страница 26: ...erval See Measurement Operation for more information Operating Temperature Range 40 to 75 C 40 to 167 F Environmental Rating Sensor and Cable Jacket Weatherproof Housing Three cup polycarbonate anemom...

Страница 27: ...place 5 Adjust the height of the sensor in the cross arm as necessary using one of the following methods and then tighten the nut and bolt until the cross arm just starts to deform a Loosen the tri cl...

Страница 28: ...ogging at a 1 second interval the sensor will report the same wind speed its calculated average for three samples before calculating and reporting a new value for another three samples Gust speed is t...

Страница 29: ...rproof Service Life 4 to 6 years typical depending upon environmental conditions Housing Injection molded plastic housing and vane static dissipating base lead free silicon bronze nose and aluminum mo...

Страница 30: ...until snug 2 Insert the sensor mounting rod into the cross arm allowing the mounting rod to protrude 1 3 cm 1 2 inch from the bottom of the cross arm 3 Tighten the nut and bolt until the rod is sligh...

Страница 31: ...urself so that the station is that number of degrees to the west of magnetic north 4 While viewing the sensor through binoculars instruct another person to rotate the sensor mounting rod to point the...

Страница 32: ...logging at a 1 second interval the sensor will report the same wind direction its calculated average for three samples before calculating and reporting a new value for another three samples Measuremen...

Страница 33: ...clamp unlimited with screws Compatible Sensors S THB 00x S TMB M0xx TMCx HB TMCx HD and external sensors from U23 00x or MX230x Maximum Sensor Diameter 1 22 cm 0 48 inches Mount temperature or temp R...