Enter the Programming Mode - HART
Press the F1 or the F2 key repeatedly, in the normal running mode, until the following
screen is shown. This enters the programming mode:
F1 F2 F3 F4
Press
YES (F4)
and then the following screen will prompt the user to enter the
password if enabled:
F1 F2 F3 F4
Enter the correct password. Default password for Level 1 is 1234.
Communication Protocol and Parameters
To program the communication parameters, press
I/O (F1)
key from the main menu.
F1 F2 F3 F4
This is the main menu for the programming mode. To exit the programming mode,
press
EXIT (F4)
repeatedly until “Normal Mode” is seen briefly. Choose I/O (F1) to
access the communication output.
F1 F2 F3 F4
Then press
COM (F1)
to select communication parameters
Set Bus protocol for HART:
F1 F2 F3 F4
Press
NXT (F1)
until HART is selected as shown and then press
OK (F4)
to accept the
setting.
HART Programming
PASWD:_
UP DN NXT OK
MAIN MENU
I/O FLO DSP EXIT
Comm=HART
NXT OK
SET PARAMETERS?
NO
YES
SET I/O
COM PUL 420 EXIT
Programming
HART