Non-Resettable
Totalizer
Reset CRC
Reset CRC
If the NVRAM CRC check fails (Error Code 36), the programmed settings values will
need to be verified and corrected before clearing the error. Call ONICON Customer
Service if you need assistance.
F1 F2 F3 F4
Press
YES (F1)
ONLY
if you want to reset the CRC and generate a new CRC value.
Non-Resettable Totalizer Activation
Regulations in some geographic locations require that flow totalizers be non-resettable.
The F-5500 can conform to these regulations.
WARNING! Once the non-resettable totalizer (NRT) has been activated on an F-5500
flow meter, the change cannot be undone. The non-resettable totalizer is only
recommended for applications that require it.
After it has been enabled, your F-5500’s totalizer and elapsed time counters will be
non-resettable.
F1 F2 F3 F4
Press
YES (F1)
ONLY if you want to set the NRT.
F1 F2 F3 F4
If you are certain you want to activate the Non-resettable totalizer, select
YES (F1)
.
11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax +1 (727) 442-5699 • [email protected]
F-5500 Thermal Mass Flow Meter Manual 06/17 - 2030 / 107023
Page 55
Model F-5500
OPERA
TION
Operation: Programming
SET NRT?
YES NO
RESET CRC?
YES NO
ARE YOU SURE?
YES NO