93
Troubleshooting your system
8
What it means
Meter is too cold (below 50°F)
to perform a test.
What to do
Move the meter and test strips to a warmer area. Insert a
new test strip when the meter and test strips are within
the operating range (50-104°F). If you do not get another
LO.t
message, you can proceed with testing.
If this message continues to appear, contact Customer
Service. Contact us at 1 888 567-3003.