100
9
Detailed information
about your system
Comparing meter results to laboratory
results
Results obtained from the OneTouch Verio Flex™ Meter
and laboratory tests are reported in plasma-equivalent
units. However, your meter result may differ from your
lab result due to normal variation. A result from your
OneTouch Verio Flex™ Meter is considered accurate when
it is within 15 mg/dL of a laboratory method when the
glucose concentration is lower than 75 mg/dL and within
15% of laboratory method when the glucose concentration
is 75 mg/dL or higher.