Detailed information about your system
9
Regression Statistics
Samples were tested on each of three test strip lots.
Results indicate that the OneTouch Verio Flex™ System
compares well with a laboratory method.
# of Subjects
# of Tests
Slope
Intercept
(mg/dL)
172
172
0.99
2.98
95% CI Slope
95% CI
Intercept
(mg/dL)
Std. Error
(S
y.x
)
(mg/dL)
R
2
0.97 to 1.01
-1.59 to 7.55
14.0
0.98