109
Detailed information about your system
9
Precision
Within Run Precision (300 Venous Blood Samples Tested
per Glucose Level)
Data generated using the OneTouch Verio Flex™ Meter.
Target
Glucose
(mg/dL)
Mean
Glucose
(mg/dL)
Standard
Deviation
(mg/dL)
Coefficient
of Variation
(%)
40
36.5
0.83
2.26
90
88.5
1.72
1.95
130
127.9
2.50
1.95
200
199.3
3.94
1.98
350
344.6
6.18
1.79
Results show that the greatest variability observed
between test strips when tested with blood is 0.83 mg/dL
SD or less at glucose levels less than 75 mg/dL or 1.98% CV
or less at glucose levels at 75 mg/dL or above.