80
Control solution testing
5
• When you first open a new vial of control solution,
record the discard date on the vial label. Refer to the
control solution insert or vial label for instructions on
determining the discard date.
• Tightly close the cap on the control solution vial
immediately after use to avoid contamination or damage.
• Control solution tests must be done at room temperature
(20–25°C). Make sure your meter, test strips, and control
solution are at room temperature before testing.
m
CAUTION
•
Do Not
swallow or ingest control solution.
•
Do Not
apply control solution to the skin or eyes as it
may cause irritation.
•
Do Not
use control solution after the expiry date (printed
on the vial label) or the discard date, whichever comes
first, or your results may be inaccurate.