117
Detailed information about your system
9
System Accuracy Results for
Glucose Concentrations <4.4 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within
±0.3 mmol/L
Within
±0.6 mmol/L
Within
±0.7 mmol/L
Within
±0.8 mmol/L
78.3%
(94/120)
99.2%
(119/120)
99.2%
(119/120)
100%
(120/120)
System Accuracy Results for
Glucose Concentrations >4.4 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within
±5%
Within
±10%
Within
±15%
Within
±20%
46.7%
(224/480)
85.4%
(410/480)
99.0%
(475/480)
99.8%
(479/480)