Troubleshooting
8
What it means
Meter is too hot (above 44°C) to work correctly.
What to do
Move the meter and test strips to a cooler area. Insert
a new test strip when the meter and test strips are
within the operating range (6–44°C). If you do not get
another Temperature too high message, you can proceed
with testing.
Temperature too high.
Out of operating range.
See Owner’s Booklet.
Warning