System Accuracy
Analytical Performance Characteristics
Diabetes experts have suggested that glucose meters should
agree within 0.8 mmol/L of a laboratory method when the
glucose concentration is lower than 4.2 mmol/L, and within
20% of a laboratory method when the glucose concentration
is 4.2 mmol/L or higher. However, experts have suggested
greater accuracy of Blood Glucose Monitoring Systems
is desirable. The OneTouch® Verio™ Pro Blood Glucose
Monitoring System has demonstrated accuracy within
±0.7 mmol/L of the laboratory method for glucose
concentrations lower than 4.4 mmol/L, and within ±15%
of the laboratory method for glucose concentrations
4.4 mmol/L or higher. Samples from 100 patients were
tested using both the OneTouch® Verio™ Pro System and
the YSI 2300 glucose analyser laboratory instrument.