SPECIFICATIONS

Dimensions

Outside    

 

29.5” wide x 23” deep x 23.6” high (750 x 584 x 600 mm) 

Weight    

110 lbs (50 kg) 

Functional

Operating Power / Current  80W / 667 mA 

 

Standby Power / Current 

7W / 100 mA

Input Voltage 

 

110-125VAC, 50/60 Hz (voltage fluctuation ±10%)

Power Connection   

Connect to branch circuit rated 20A max

Fuse Protection    

10 A at 250V

    

Environmental

Operating  

65°F (18 °C) to 105 °F (40.5 °C) temperature

  10–95% 

humidity

  

Note: 45°F (7 °C) to 65°F (18 °C) temperature must run warm-up  

 

 

cycle before disinfecting. See Operating. 

Storage    

-20 °F (-28 °C) to 140 °F (60 °C) temperature

  10–95% 

humidity

  

Note: Allow unit to reach Operating temperature before use. 

Pollution Degree  2

UV-C Energy Source

Germicidal Bulbs   (4) 20W UV-C Lamps
Output    

254 nm UV-C Light

Chamber Components

Controller  

See Touchscreen Display. 

Emergency Switch  

Located on both sides on machine, buttons turn OFF chamber 
in case of emergency (fire, smoke, etc.).

Tray / Reflective Liner  The tray / reflective liner rests inside the bottom of the chamber 

and are removable for easy cleaning. 

Handles 

Can be found on either side to securely open and close both 
chamber doors.

Windows 

The windows use glass not transparent to UV-C.

Germicidal Bulbs  

There are four (4) germicidal UV-C 20W bulbs.

Chamber 

Chamber will automatically turn off after programmed time.

Magnet  

Securely holds doors in closed position while running.

3

Содержание BioShift

Страница 1: ...user guidelines and additional operation instructions for agricultural industrial use For Service or Parts Call Your Sales Representative or Contact us at T 1 763 381 5621 F 1 763 381 5698 E info onc...

Страница 2: ...nents 7 Maintenance Screen Components 8 Settings Screen Components 9 Programming 10 Cycle Time 10 Sleep Timer 10 Maintenance 11 13 Alert Notifications Errors 11 Check Lamp Life 12 Order Replacement La...

Страница 3: ...biocidal and virucidal effectiveness Note UV C radiation is only effective on hard surface objects It does not penetrate through materials and is unable to kill bacteria viruses and pathogens on shro...

Страница 4: ...Allow unit to reach Operating temperature before use Pollution Degree 2 UV C Energy Source Germicidal Bulbs 4 20W UV C Lamps Output 254 nm UV C Light Chamber Components Controller See Touchscreen Dis...

Страница 5: ...o the eyes and skin Never look into an illuminated UV C lamp without proper eye protection The chamber windows provide a safe way to view inside chamber without the risk of UV C expsosure Ozone may be...

Страница 6: ...e door has been opened the clean side door will not open unless another cycle is complete The BioShift chamber WILL NOT resume operation after re closing Pressing the EMERGENCY STOP will open clean si...

Страница 7: ...ng bracket located on both sides of unit and set aside for later use 2 Using screws or nails not provided attach removed panels to support beams Use as many screws or nails needed to ensure panel is s...

Страница 8: ...DISPLAY The touchscreen display is located on the clean side of the BioShift chamber Home Screen Components 1 Disinfection Time Display 2 Chamber Start Stop 3 Chamber Status 4 Settings 5 Maintenance...

Страница 9: ...under the Maintenance button Maintenance Screen Components 1 View Lamp Life Remaining 2 Reset Lamp Life 3 View Lamp Replacement Info 4 Run 100mJ cm2 Dose Test 5 Run 250mJ cm2 Dose Test 6 Run 1000mJ c...

Страница 10: ...9 1 4 2 3 TOUCHSCREEN DISPLAY Located under the Home screen Settings button Settings Screen Components 1 View Set Lamp Cycle 2 Set Lamp Cycle 3 Set Sleep Time 4 Expired Lamp Life 5 Contact ONCE 5...

Страница 11: ...MM SS Once desired lamp life cycle has been chosen select the check mark to set PROGRAMMING ITEMS TO NOTE ControllercannotbeprogrammedifEMERGENCYSTOPbuttonsarepressedin Pressing EMERGENCY STOP button...

Страница 12: ...pear in the upper left corner of Maintenance button when the following occurs UV C lamps have reached maximum cycle hours Replacement bulbs have been installed and Lamp Life needs to be reset Simply f...

Страница 13: ...U 26 0053 110 240V Replacement Ballast SKU 26 0052 Changing Lamps and Ballasts To replace lamps simply twist the lamp and pull out from its housing Slide the new lamp in and twist to lock it in place...

Страница 14: ...d a soft towel Be careful not to spray or apply chemicals directly on the UV C lamps This may result in total lamp destruction UVC Dose Verification It is imporatant to regularly verify the UVC dose o...

Страница 15: ...G FUNCTIONS BioShift chamber doors shall not be opened during radiation cycle BioShift chamber doors cannot be open at the same time under normal operation In case of emergency EMERGENCY STOP button c...

Страница 16: ...caused by accident or misuse 3 Damage caused by disaster or other acts of God 4 Corrosion due to use of improper solvents or detergents 5 Repair or modification performed by anyone other than ONCE or...

Страница 17: ...s s par une catastrophe ou d autres v nements de force majeure 4 Corrosion due l utilisation de solvants ou de d tergents inappropri s 5 R paration ou modification effectu e par une personne autre que...

Страница 18: ...use and documented cleaning record Upon request ONCE representatives shall be granted access to the defective Product system or application for verification of non compliance If requested by ONCE the...

Страница 19: ...express or implied including without limitation any warranty of merchantability or fitness for a particular purpose but can make available to the Purchaser upon request but only to the extent permitt...

Страница 20: ...ave entered into with any party in relation to the supply provision donation and or otherwise making available of BioShift Chambers Signify explicitly excludes any and all liability towards any party...

Страница 21: ...de mod le du produit d faillant 2 Date d installation et informations sur la facture 3 Description d taill e du probl me notamment syst mes lectriques pertinents emplacement de l installation et symp...

Страница 22: ...e limit e la p riode de garantie est de deux 2 ans compter de la date de livraison sous r serve d un nettoyage document des produits au moins une fois tous les trois 3 mois au cours de la p riode de d...

Страница 23: ...s masques chirurgicaux et ou tout autre EPI de qualit pour dispositifs m dicaux Signify d cline toute responsabilit envers toute partie si des chambres BioShift sont utilis es pour d sinfecter les sur...

Страница 24: ...lid for all 24 0200 models For Service or Parts Call Your Sales Representative or Contact us at T 1 763 381 5621 F 1 763 381 5698 E info once group www once group Once Inc a Signify company 15255 23rd...

Отзывы: