MAINTENANCE

Reset Lamp Life

• 

From the Home screen, choose the Maintenance button (see Maintenance Screen 
Components).

• 

Choose the RESET LAMP button. 

• 

Lamp Life will be reset to default, unless custom values have been entered in the 
Lamp Life Settings menu.

Cleaning 

• 

Do not use industrial degreasers or strong solvents such as brake cleaners.

• 

Do not override safety features during maintenance / cleaning.

• 

When cleaning the reflective tray liner, do not use abrasives, they will reduce the 
reflectivity. For difficult to clean areas, soak the entire liner in a mild detergent. 

• 

Outside: BioShift

®

 unit can be cleaned with a damp towel. Strong solvents or 

detergents are not recommended. A typical household cleaning solvent is 
recommended.

• 

Inside: Clean inside chamber walls with ethanol (reagent grade) and a soft towel. 
Be careful not to spray or apply chemicals directly on the UV-C lamps. This may 
result in total lamp destruction.

 

UVC Dose Verification

It is imporatant to regularly verify the UVC dose output of the BioShift unit to ensure 
proper disinfection. Every 50 cycles or once per month, whichever comes first, the 
operater shall verify that the UVC dose exceeds a minimum of 100 mJ/cm

2

 in a short 

programmed test cycle. With an empty chamber, place a card (below) at the center of 
the shelf face-up. On the touchscreen display, select the “100 mJ/cm

Dose Test” button 

to start a test cycle. Upon completion of the test cycle, the operator will evaluate the 
color development of the center circle. If the center color matches or exceeds the 100 
mJ/cm

region, the unit has passed the test. If the color doesn’t match or exceed 100 mJ/

cm

2

, all the lamps in the unit must be replaced. 100 mJ/cm

2  

test cards are available from 

Once. 

Note: Do not use dose verification as a disinfection cycle.

 

13

Содержание BioShift

Страница 1: ...user guidelines and additional operation instructions for agricultural industrial use For Service or Parts Call Your Sales Representative or Contact us at T 1 763 381 5621 F 1 763 381 5698 E info onc...

Страница 2: ...nents 7 Maintenance Screen Components 8 Settings Screen Components 9 Programming 10 Cycle Time 10 Sleep Timer 10 Maintenance 11 13 Alert Notifications Errors 11 Check Lamp Life 12 Order Replacement La...

Страница 3: ...biocidal and virucidal effectiveness Note UV C radiation is only effective on hard surface objects It does not penetrate through materials and is unable to kill bacteria viruses and pathogens on shro...

Страница 4: ...Allow unit to reach Operating temperature before use Pollution Degree 2 UV C Energy Source Germicidal Bulbs 4 20W UV C Lamps Output 254 nm UV C Light Chamber Components Controller See Touchscreen Dis...

Страница 5: ...o the eyes and skin Never look into an illuminated UV C lamp without proper eye protection The chamber windows provide a safe way to view inside chamber without the risk of UV C expsosure Ozone may be...

Страница 6: ...e door has been opened the clean side door will not open unless another cycle is complete The BioShift chamber WILL NOT resume operation after re closing Pressing the EMERGENCY STOP will open clean si...

Страница 7: ...ng bracket located on both sides of unit and set aside for later use 2 Using screws or nails not provided attach removed panels to support beams Use as many screws or nails needed to ensure panel is s...

Страница 8: ...DISPLAY The touchscreen display is located on the clean side of the BioShift chamber Home Screen Components 1 Disinfection Time Display 2 Chamber Start Stop 3 Chamber Status 4 Settings 5 Maintenance...

Страница 9: ...under the Maintenance button Maintenance Screen Components 1 View Lamp Life Remaining 2 Reset Lamp Life 3 View Lamp Replacement Info 4 Run 100mJ cm2 Dose Test 5 Run 250mJ cm2 Dose Test 6 Run 1000mJ c...

Страница 10: ...9 1 4 2 3 TOUCHSCREEN DISPLAY Located under the Home screen Settings button Settings Screen Components 1 View Set Lamp Cycle 2 Set Lamp Cycle 3 Set Sleep Time 4 Expired Lamp Life 5 Contact ONCE 5...

Страница 11: ...MM SS Once desired lamp life cycle has been chosen select the check mark to set PROGRAMMING ITEMS TO NOTE ControllercannotbeprogrammedifEMERGENCYSTOPbuttonsarepressedin Pressing EMERGENCY STOP button...

Страница 12: ...pear in the upper left corner of Maintenance button when the following occurs UV C lamps have reached maximum cycle hours Replacement bulbs have been installed and Lamp Life needs to be reset Simply f...

Страница 13: ...U 26 0053 110 240V Replacement Ballast SKU 26 0052 Changing Lamps and Ballasts To replace lamps simply twist the lamp and pull out from its housing Slide the new lamp in and twist to lock it in place...

Страница 14: ...d a soft towel Be careful not to spray or apply chemicals directly on the UV C lamps This may result in total lamp destruction UVC Dose Verification It is imporatant to regularly verify the UVC dose o...

Страница 15: ...G FUNCTIONS BioShift chamber doors shall not be opened during radiation cycle BioShift chamber doors cannot be open at the same time under normal operation In case of emergency EMERGENCY STOP button c...

Страница 16: ...caused by accident or misuse 3 Damage caused by disaster or other acts of God 4 Corrosion due to use of improper solvents or detergents 5 Repair or modification performed by anyone other than ONCE or...

Страница 17: ...s s par une catastrophe ou d autres v nements de force majeure 4 Corrosion due l utilisation de solvants ou de d tergents inappropri s 5 R paration ou modification effectu e par une personne autre que...

Страница 18: ...use and documented cleaning record Upon request ONCE representatives shall be granted access to the defective Product system or application for verification of non compliance If requested by ONCE the...

Страница 19: ...express or implied including without limitation any warranty of merchantability or fitness for a particular purpose but can make available to the Purchaser upon request but only to the extent permitt...

Страница 20: ...ave entered into with any party in relation to the supply provision donation and or otherwise making available of BioShift Chambers Signify explicitly excludes any and all liability towards any party...

Страница 21: ...de mod le du produit d faillant 2 Date d installation et informations sur la facture 3 Description d taill e du probl me notamment syst mes lectriques pertinents emplacement de l installation et symp...

Страница 22: ...e limit e la p riode de garantie est de deux 2 ans compter de la date de livraison sous r serve d un nettoyage document des produits au moins une fois tous les trois 3 mois au cours de la p riode de d...

Страница 23: ...s masques chirurgicaux et ou tout autre EPI de qualit pour dispositifs m dicaux Signify d cline toute responsabilit envers toute partie si des chambres BioShift sont utilis es pour d sinfecter les sur...

Страница 24: ...lid for all 24 0200 models For Service or Parts Call Your Sales Representative or Contact us at T 1 763 381 5621 F 1 763 381 5698 E info once group www once group Once Inc a Signify company 15255 23rd...

Отзывы: