Electromagnetic Immunity
GUIDANCE AND MANUFACTURER’S DECLARATIONELECTROMAGNETIC IMMUNITY
The Laser is intended for use in a typical commercial or hospital environment. The customer or
the user of the Laser should assure that it is used in such an environment
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2 ed2.0 (2008-12)
± 8 kV contact
± 15 kV air
complies with IEC 60601
floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4 ed3.0 (2012-04)
± 2 kV for AC mains
power lines
±1 kV for signal input/
output lines
complies with IEC 60601
Mains power quality should be that of a typical
commercial or hospital environment.
Surges Immunity
IEC 61000-4-5 ed3.0 (2014-05)
± 2 kV
complies with IEC 60601
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
ed2.0 (2004-03)
0 % UT
(100% dip in UT) for 0.5 cycle
0 % UT (100 % dip in
UT) for 1 cycle
70 % UT (30 % dip in
UT) for 25 cycles
110VAC/60Hz
240VAC/50Hz
complies with IEC 60601
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the laser requires continued
operation during power mains interruptions, it is
recommended that the Laser be powered from
an uninterruptible power supply or a battery.
Power frequency magnetic field
IEC 61000-4-8 ed2.0 (2009-09)
30 A/m
50/60 Hz
complies with IEC 60601
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: U
T
is the AC main voltage prior to the application of the test level
Electromagnetic Immunity II
Immunity test
IEC 60601-test level
Compliance level
Radiated, Radio-frequency,
Electromagnetic Immunity
IEC 61000-4-3 ed3.0 (with A1:2007+A2:2010)
3 V/m V
eff
80 MHz to 6000 MHz
3 V/m V
eff
conforms to IEC 60601
Conducted, Radio-frequency,
Electromagnetic Immunity Test
IEC 61000-4-6 ed2.0 (with A1:2004+A2:2006)
3V
rms
/ 6V
rms
150 kHz to 80 MHz
3V
rms
/ 6V
rms
conforms to IEC 60601