L-1670 Rev 0
45
© OmniGuide, Inc.
5.
Do you hear the laser firing tone when the foot pedal is pressed?
If the problem persists, contact OmniGuide for assistance
Low power or no power is observed out of the OmniGuide Fiber:
1.
Prior to surgery the laser is to be test fired with a fiber to confirm power
output. If no energy is emitted from a fiber make sure the fiber is not clogged.
Clean in sterile solution if necessary. Replace fiber if necessary.
2. Verify that the Laser is on and that it is set to “Ready”.
3. Verify that gas of the proper type and pressure is being supplied to the gas hose of the Adapter.
The proper gas type and pressure is shown on the insert provided with each fiber.
4. Verify that the Adapter is securely mounted on the Laser. If the Adapter exhibits any up-and-down, rocking,
or rotational movement relative to the Laser, discontinue use and contact OmniGuide for assistance.
5.
Set the Laser power to 4W in CW mode. Keep the cannula of the handpiece (if present)
and the fiber as straight as possible and operate Laser for up to 5 seconds while the
fiber’s distal-end is 3mm from a moist tongue depressor. Use aseptic technique. If no
visible burn spot appears, replace the OmniGuide fiber and repeat this test.
If the problem persists, contact OmniGuide for assistance
System too hot
1. Is the laser system wrapped too tightly in sterile wrap or the air inlet on the right side is blocked?
2.
Is there restriction in the air flow at the bottom of the system?
3.
Is the room temperature below 30° C (86F)?
4. Have you treated patients with full power more than 30 minutes?
5. Is the system close to any heat source, for example room heating?
If the problem persists, contact OmniGuide for assistance. Call OmniGuide Service at 888-666-4484 ext. 3.
16.0 WARRANTY
Point of Sale Warranty
for point of sale warranty on new and refurbished units please refer to OmniGuide’s terms and conditions.
Parts and Service Warranty
OmniGuide warrants that Parts installed and services performed by OmniGuide or OmniGuide-
approved subcontractors, will be free from defects in material and workmanship respectively
for a period of 90 days from the date of installation or performance of the service, respectively,
provided that the warranty for parts purchased directly by the Customer from OmniGuide
and not installed by OmniGuide is 30 days only from the date of purchase.
Services Remedy.
If OmniGuide determines that any Service fail to meet the foregoing warranty, OmniGuide
shall re-perform such Service to meet the foregoing warranty at its cost.