Functionality
The display screen, RfID antenna, articulating arm, wheels, emergency machine
off, rear panel and foot switch are to be checked for routine functionality. The
unit is to be inspected for any loose, missing or damaged hardware.
Parts
The BEACON Advanced Energy Laser System cart, and accessories are to be inspected for
any loose, damaged, or missing parts. The following parts are to be replaced annually:
14.1.1 Fan filter element (located under the air intake enclosure panel
on the right side of the laser), PN 1-0092-015-00-00
14.1.2 Yoke washers for the CGA-930 fitting that connects the inlet hoses for
the gas management assembly to the tanks, PN 1-0090-017-00-00
OmniGuide Service
for additional information on the Yearly Preventive Maintenance please
contact OmniGuide Service at 1-888-666-4484.
15.0 TROUBLESHOOTING
System does not start
1. Check if electrical power cord is fully securely plugged into an individual outlet
2. Check if electrical power cord is fully plugged into the Laser
3. Check if mains switch at the laser base is in “ON“ position
4. Check if emergency switch is pulled out all the way
5. Check if the door interlock connector is in place at base of Laser
6. Check fuses at the mains connector at base of the Laser. If the fuses are
blown, exchange fuses with spare fuses supplied with the system.
7. Door is interlocked, and the door is open.
If the problem persists, contact OmniGuide service for assistance (1-888-666-4484)
Password does not work
Retype the pin code (password). If you cannot get into the main menu, restart the system and retry. If you
have no success entering the main menu contact OmniGuide service for assistance (1-888-666-4484).
No effect on tissue
1.
Prior to surgery the laser is to be test fired to confirm the aiming beam and CO beam are aligned
and that there is no clipping or observed disfigured beam if the articulating arm is to be used.
2. Increase the power of the system
3. Is the “Ready” button pressed?
4. Increase the pulse length