12.1.17 From the user graphical interface set the appropriate
Pulse Mode, Helium Flow, and Power Level
12.1.17.1 Surgeon custom laser settings can be saved and recalled using “Save User Settings”
and “Load User Settings” graphical interface features accessibly in the “Settings” menu
12.1.18 Provide the foot pedal to the surgeon
12.1.19 Place the laser in ready mode and press the foot pedal to activate the laser
12.1.19.1 It is strongly recommended the fiber be dipped in sterile saline to confirm gas flow and
a test shot performed on a sterile tongue blade to confirm proper power setting prior to use
12.1.19.2 The fiber should be cleaned routinely during surgery
to prevent tissue build up on the distal end
12.1.20 Power, gas, and pulse settings should be adjusted accordingly
throughout the procedure at the discretion of the attending surgeon
12.1.21 Laser modality can be changed at any time by pressing the
fiber icon in the upper left and choosing “change mode.”
Arm Usage Mode
12.1.22 Enter the password and review any notifications from system start-up
12.1.23 Select the “arm” icon
12.1.24 Determine what accessories will be used with the articulating arm
12.1.24.1 Consult section 5.4. of the current manual for the list of compatible accessories.
12.1.25 At the user screen set the Lasing Mode, Aiming beam brightness, and Power Setting
12.1.26 Position the laser so that mobility of the arm will not conflict or
interfere with other equipment and will not pull on the laser body
12.1.27 Provide the foot pedal to the surgeon
12.1.28 Place the laser in ready mode and press the foot pedal to activate the laser
12.1.28.1 It is strongly recommended a test shot be performed on a sterile
tongue blade prior to use to confirm the aiming beam and CO
2
beam are
mutually aligned and that the power setting is appropriate
12.1.29 Power and pulse settings should be adjusted accordingly throughout
the procedure at the discretion of the attending surgeon
12.1.30 Laser modality can be changed at any time by pressing the
“Arm” icon in the upper left and choosing “change mode.
Post Procedure Breakdown
12.1.31 Press the power button on the front of the laser to turn it off
12.1.32 When the unit has powered down it can be unplugged from the wall.