Specifications 

Performance 

Input Type: 

Strain Gauge Full Bridge Sensors 

 

Input Range: 

Up ±5mV/V (±50mV/V can be supplied, with factory set option) 

 

Non Linearity: 

±0.005% FSD 

 

Thermal Drift: 

<5 ppm/°C 

 

Excitation Voltage: 

5Vdc (±4%), 59mA maximum current  

 

Minimum Bridge Resistance:  85

 (4off 350

 sensors in parallel) 

 

Internal Battery: 

2off AA size alkaline, access via sealed rear compartment 

 

Battery Life: 

45 hours (Typical 450 hours in low power mode), with 350

 sensor 

 

Update Rate: 

Up to 40mS (can be set in configuration menu) 

Indication 

Display Type: 

7 digit LCD display, 8.8mm high digits 

 

Display Resolution: 

1 part in 250,000 at 1Hz update rate 
1 part in 65,000 at 10Hz update rate 

 

Annunciators: 

Low Battery warning; peak; trough; hold; net; shunt cal; range 

Control Variables 

Front Panel User Keys: 

Tactile Keys with recessed rims for:- 
ON/OFF 

Switches TR150 power on/off 

RANGE 

Selects between two ranges 

HOLD 

Hold the current display value, press again to release

GROSS/NET  Zero’s display (±100% range) 
SHUNT CAL  Generates simulated input for indicator testing 
PEAK Enables 

peak 

hold 

TROUGH 

Enables valley/trough hold 

 

Settable Parameters: 

Tare/Zero value; display resolution/decimal point position; display 
update rate; low power mode; auto power off;  

Mechanical 

Electrical Connection: 

5 pin Binder socket (mating plug supplied) 

 

Physical Size: 

See drawing below 

 

Weight: 260 

grams 

 

Legends: 

Insert legends for engineering unit identification (supplied) 

Environmental 

Operating Temperature: 

-10°C to +50°C 

 

Environmental Rating: 

IP65 (when mating plug fitted) 

 

Enclosure Type: 

ABS, dark grey (Leather Carry Case Optional) 

 

Safety/Low Voltage Directive 

73/23/EEC amended by 93/68/EEC To IEC 1010-1:1990, EN 
61010 – 1 – 1993 89/336/EEC 

 

EMC Directive 

EN 50 081 – 1 : 1992 (Light Industrial) 

 

Emissions 
 

EN 50 081 – 2 : 1992 (Heavy Industrial) 
pr EN 50 093 : 1991 

 

EMC Emissions 

 

EN 50 082 – 1 :1992 (Light Industrial) 
EN 50 082 – 2 :1992 (Heavy Industrial) 

 

Mechanical Dimensions

 

 

 

 

Lbs

kN

kgf

90

152

34

UK / Europe Office 
Tel: +44 (0)8700 434040  
Fax: +44 (0)8700 434045   

[email protected]  
www.omniinstruments.co.uk

 

Australia / Asia Pacific Office 
Tel +61 (0)282 442 363 
Fax +61 (0)294 751 278

 

[email protected]

 

www.omniinstruments.com.au 

USA / Canada Office 
Tel +1-866-849-3441 
Fax +1-866-628-8055

 

[email protected]

 

www.omniinstruments.net

Содержание TR150

Страница 1: ...44 0 8700 434045 info omniinstruments co uk www omniinstruments co uk Australia Asia Pacific Office Tel 61 0 282 442 363 Fax 61 0 294 751 278 info omniinstruments com au www omniinstruments com au USA...

Страница 2: ...11 SHUNT CAL Button 11 PEAK Button 11 TROUGH Button 11 CONFIGURATION MENU Parameters 12 CALIBRATION MENU Parameters 14 Calibration Procedures 15 Specifications 17 Mechanical Dimensions 17 UK Europe O...

Страница 3: ...TR150 is powered by two internal non rechargeable AA alkaline batteries There is an option to have rechargeable alkaline batteries which can be charged without the need for removing them from the TR1...

Страница 4: ...PIN 4 ve Signal PIN 5 ve Charge Input RS232 Port Connections If the TR150 has been ordered with the optional RS232 output then this will be available via an 8 pin 723 series Binder connector The wiri...

Страница 5: ...ctions If this is performed at the end of the calibration procedure then a figure can be noted so the user can check calibration accuracy or connection integrity The button has to be held down to oper...

Страница 6: ...Et ZErO 0000000 SEt rAtE SEt OUEr SEt OPEr AutO OFF rS232 25 10 3 1 0 5 0000000 PSAVE 00 EnAbLEd RETURN TO NORMAL DISPLAY MODE UK Europe Office Tel 44 0 8700 434040 Fax 44 0 8700 434045 info omniinstr...

Страница 7: ...000000 dISP HI 0000000 donE tAbLE InPut LO 0000000 dISP LO 0000000 InPut HI 0000000 dISP HI 0000000 donE CAL VAL SEt 9Ain 0000000 SEt OFFS donE 0000000 UK Europe Office Tel 44 0 8700 434040 Fax 44 0 8...

Страница 8: ...mode When you choose to change the update rate you will be prompted whether you want to select 25Hz If you do not press You will then be prompted to select any of the other values which in order are 1...

Страница 9: ...tons to select a digit and the and buttons to increment or decrement the digits Press to accept the value and move onto the next parameter rS232 PRESS TO SKIP THIS PARAMETER AND EXIT MENU PRESS TO ENA...

Страница 10: ...ps of 0 005 The decimal point position is moved one place to the right each time you press the and together Any value can be entered for the resolution using the and buttons to select a digit and the...

Страница 11: ...down the button for 3 seconds It is also possible to set an Auto off value in the configuration menu so that the TR150 automatically switches itself off if there is no keyboard activity within a prese...

Страница 12: ...an idea of the calibration accuracy at a later date or for checking the integrity of the sensor and sensor cabling The shunt calibration resistor can be changed to suit specific requirements It is su...

Страница 13: ...in the display To remove the alarm the display value must be reduced to a value that is lower than that set in the SEt OVEr parameter This can be very useful as a safety feature or simply as a quick...

Страница 14: ...detected during the 25 minutes then the time period is restarted This can be a useful feature in a site environment should the TR150 be left unintentionally powered on rS232 Parameter This parameter...

Страница 15: ...e decimal point position one place to the right It also allows for the setting of the display resolution or the number of display counts the display changes with an input change To change the resoluti...

Страница 16: ...w stimulus applied to the sensor The value can be entered by using the and buttons to select a digit and the and buttons to change the digit When the value has been set press 6 You will then be prompt...

Страница 17: ...e and buttons to change the digit When the value has been set press 9 You will then be prompted with dISP HI Press to enter the display value required for the high input figure entered The value can b...

Страница 18: ...PEAK Enables peak hold TROUGH Enables valley trough hold Settable Parameters Tare Zero value display resolution decimal point position display update rate low power mode auto power off Mechanical Ele...

Отзывы: