16
tAbLE
Calibration Procedure
(1) When
CALibrAt
is displayed, press
.
(2)
LiVE ?
will now be displayed. Press
.
(3)
tAbLE ?
will now be displayed. Press
.
(4) You will be prompted with
InPut LO
. Press
.
(5) Now enter the zero offset sensitivity of the sensor by using the
and
buttons
to select a digit and the
and
buttons to change the digit. When the value has
been set, press
. If you do not know this, simply enter all zeros.
(6) You will then be prompted with
dISP LO
. Press
to enter the display value
required for the low input figure entered. The value can be entered by using the
and
buttons to select a digit and the
and
buttons to change the digit.
When the value has been set, press
.
(7) You will then be prompted with
InPut HI
. Press
.
(8) Now enter the sensitivity figure supplied for the sensor by using the
and
buttons to select a digit and the
and
buttons to change the digit. When the
value has been set, press
.
(9) You will then be prompted with
dISP HI
. Press
to enter the display value
required for the high input figure entered. The value can be entered by using the
and
buttons to select a digit and the
and
buttons to change the digit.
When the value has been set, press
.
(10) You should now see
donE
displayed. This means the calibration was successful.
Press
to return the TR150 to normal operation mode, with the new calibration
data stored. If you see
FaiLEd
, then you will need to repeat the calibration,
checking that you have completed the procedure in the correct order, and that the
sensor is connected correctly.