6
Troubleshooting
TICRO 100
automatically powers up after supply voltage is provided via one of the DC power inputs or
via Ethernet (PoE). The green LED S1 lights up and the red LED S2 is off when the device is ready for
operation. In case of operational problems with
TICRO 100
, please use the following table to solve the
problem.
The TICRO 100 User Manual contains detailed descriptions of the operating procedures possible for
TICRO 100
directly on the device and via its Web Interface.
Reset pushbutton
LEDs
S1 & S2
LEDs
RJ45 & FIBER
6.1
Troubleshooting procedures
Problem / LED Indication Description / Possible Cause / Remedy
Green LED S1 is on
Red LED S2 is off
TICRO 100
is ready for operation.
LEDs S1 and S2 are
continuously off
TICRO 100
is not supplied with power.
TICRO 100
may be supplied
using the following power supply options. See section
page 16 for more detailed information.
1. Power supply via Power over Ethernet (PoE).
2. Power supply via the plug-in power supply unit provided on delivery.
3. Power supply via any suitable external DC supply voltage.
Green LED S1 flashes
Red LED S2 is off
TICRO 100
is in the recovery mode, waiting for new software. Refer to
chapter "Operating the TICRO 100" of the TICRO 100 User Manual.
Troubleshooting
OMICRON
21