9.5.4.2
Digital Sensor Trigger Function
The Sensor Trigger function is used to reset the IPSO min/max values as well as control the Calibration
process.
Digital Sensor Trigger Function
7
6
5
4
3
2
1
0
0
0
0
0
0
0
0
Reset
Min/Max
15
14
13
12
11
10
9
8
0
0
Calibration
Reset
Calibration
Status
Calibration
Mode
Capture
High
Capture
Low
Calibration
Start
Setting the Reset Min/Max bit to 1 will reset the Min/Max values recorded by the IPSO process. No User
Calibration process is supported on the Digital inputs and all configuration bits should be written as 0.
9.6
Process Input
The Process Input interface provides single-ended voltage and current loop inputs. The Sensor Configuration and
Sensor Device fields may be written to provide control of the overall function of the channel and the signal types
used.
9.6.1
Process Descriptor
Offset
Name
Value
Description
0x00
Measurement
Type
0x??
Analog Voltage and Current – set by Sensor Type
field in Configuration byte.
0x01
Data Type/Format
0x06
Float
0x02
Configuration
0x4?
Determines Process input type/range
0x03
Sensor Device
0x??
Determines AIO signal types
0x04..0x08
UOMR
“??”
Units of measure
9.6.1.1
Process Measurement Types
The Process interface provides a measurement dependent on the input range/type selected. The units of
measure may be changed by the user.
Sensor
Type
SI Derived
Units
Measurement
0x11
mV
Process Voltage (0 - 1.0 V, 0 – 2.0 V)
0x13
mA
Process Current (0-24 mA current loop, common return)
9.6.1.2
Process Input Data Type/Format
Process Input Data Type/Format
7
6
5
4
3
2
1
0
Smart
Sensor Writeable
Factory
Calibrate
Reserved
Data Type
0
0
?
0
0x06 = Float
9.6.1.2.1
Data Type
The 4-bit Data Type field determines the type of data of the specific sensor (see Data Types).
30
|