47
REMOTE
Puts the calibrator in remote mode. From the remote mode the user
can still use the keypad to get back to local unless the command
LOCKOUT was entered before REMOTE. Than the keypad is totally
locked out, and the user has to send the LOCAL command to get back
to local operation.
RTD_TYPE
Sets the RTD type. The following is a list of RTD types they way they
should be entered after the command:
PT385_10;
PT385_50;
PT385_100;
PT385_200;
PT385_500;
PT385_1000; PT392_100;
PTJIS_100; Ni120;
Cu10; Cu50;
Cu100; YSI_400;
OHMS;
CUSTOM;
For example:
RTD_TYPE PT385_10 sets RTD type to Pt385-10
RTD_TYPE?
Returns the RTD type.
RTD_WIRE
Sets the number of wires used for connection in measuring RTDs. The
calibrator measures RTDs using 2-, 3-, and 4-wire connections. After
the command, enter 2W for 2- wire, 3W for 3-wire, and 4W for 4-wire.
For example:
RTD_WIRE 4W sets the connection to 4-wire
RTD_WIRE?
Returns the number of wires used in the RTD connection.
SIM
Sets the output for current simulation. This command also switches the
calibrator into mA simulation mode. A number and a unit must be
entered after the command. For example:
SIM 5 MA sets the current simulation at 5 mA