17
4.1 Setting 0% and 100% Output Parameters
To set the 0% and 100% points, use the following steps:
1. Select the lower display [LOWER] from Main Menu, and choose
the desired primary parameter.
2. Select output [OUT] from the input/output control, and enter the
desired value. For example select [VOLTS OUT].
3. Enter 5V with the keypad and press Enter.
4. Press any one of the four cursor control arrows to display the
setpoint control menu.
5. Hold down the Function Key that corresponds to 0% [F1]. 0%
will flash and the setpoint is stored.
6. Repeat these steps, entering 20V and holding the Function Key
that corresponds to 100% [F3].
7. Use the 25% key to cycle 5 V and 20 V in 25% increments.
4.1-1 Stepping the current output
To use the 25% function with mA output, follow these steps:
1.
Select the lower display from the Main Menu, and choose
mA.
2.
Use the 25% key to cycle between 4 mA and 20 mA in 25
% intervals.
4.2 Using the Automatic Output Functions
There are two automatic output functions, step and ramp. The
selected function can be turned on and off using the Output Home
Menu. The Automatic Output Function parameters can be set in the
Auto Function Menu.
Parameters include:
1. Which auto function will be available (Step or Ramp).
2. The Auto Function Time, time between steps for step and time to
get from over one limit to the next for ramp.
The limits for the ramp and step functions are set to the 0% and
100% values. See Section 4.1 Setting 0% and 100% Output
Parameters. Steps are in 25% increments from the 0% value to the
100% value.
4.3 Sourcing mA
To source a current, follow these steps:
1. From the Main Menu select lower display [LOWER]. Choose
[mA] from the primary parameters.