Configuring sensors
Enabled Units
Select or deselect measurement units to view them on the Smart
Manometer. They are saved on each sensor.
▪
The
All
button selects all measurement units.
▪
The
None
button deselects all measurement units.
▪
You must select at least one unit before you can
Write
to the
sensor.
Two User Defined Units
You can define the
Name, Multiplier, Offset,
and
Function
for two
non-standard or unsupported display units.
Note: Flow type does not use Offset.
Note: (multiplier × PSI) + Offset = the displayed value for linear
type.
Damping
No settings on the Smart Manometer
You cannot set the Damp Rate on the Smart Manometer. You have
to click the
Configuration
button in
HHP400 Suite
to change the
Damp Rate.
Damping rate
Select a
Time Constant:
1 second, 2 seconds, 4 seconds,
8 seconds, or 16 seconds.
▪
When set for 4 second time constant, it takes 4 seconds
from the time of the step change until the Smart Manometer
displays the full value of the new pressure.
Damping types
Select a
Damp Type:
Exponential, Smart Quick, Smart Normal, or
Smart Slow.
▪
Use adjustable
Exponential
type damping to steady the
display when you measure pulsating pressure or flow.
▪
Exponential
damping displays approximately 70 % of a step
change in pressure at the next display update.