M20013 REL-0000 (2103)
User Manual HHP460 Smart Manometer with Altimeter & Air Speed Test
31 of 72
1. Press the
Backward
key a second time if you do not want
change the duration and
TAP START TO BEGIN
.
(
START
is referring to the
Start
key).
2. The numbers blink to indicate when the test has stopped. This
screen displays the results of the Leak Test with the calculated
leak rate in units per minute.
3. After the test has stopped, press the
Forward
key to review
the MIN value, and then the MAX value, and then you can
select an
--out-
option to exit the test by pressing either
the
Cancel
or
Accept
keys.
Note: MIN is the minimum pressure measured during the test.
MAX is the maximum pressure measure during the test.
Note: Starting a Data Log is different than starting a Leak Test.
Data Log requires pressing and holding the
Start
key until you
see the word
Starting
.
In a Leak Test press and release the
Start
key (the on-screen message refers to this action as TAP).