Troubleshooting Guide
Symptom
Possible Cause
Method of Correction
Unit Leaks
Fittings not tight enough.
Sensor assembly cracked
Tighten fittings (see
section B4)
Unit must be returned for
repair (see Section D3)
No output signal or flow
indication
No power or low power
No flow passing through
unit
Output shorted or
overloaded
Improper connection
Moisture on internal
electronics
Output circuitry damaged
Liquid not translucent
enough.
Attempting to measure
flows below the specified
minimum for the unit
Apply correct power
Display or output should
read zero if there is no
flow. Increase the flow
rate until a reading is
obtained. (Do not exceed
the specified maximum
flow!)
Check electrical
connections (see section
B5)
Check electrical
connections (see section
B5) and ensure connector
is correctly plugged in.
Allow unit to dry out in a
dry environment.
Unit must be returned for
repair (see Section D3)
Contact the Omega
Engineering Service
Department for
assistance.
Use higher flow rates.
Flow indication “bounces”
or fluctuates
Flow is actually fluctuating
Not using the fittings
supplied by the
manuafacturer
Entrapped air or gas
Turbulence in tubing or
sensor
Fluctuating input power
Mechanical sensor
assembly damaged
Correct flow source to
obtain steady flow
Use the supplied fittings
Remove entrapped air
(see section C2)
Straighten the inlet tubing
or lengthen it to at least
6” (150mm)
Correct the power source
or change to a regulated
supply
Unit must be returned for
repair (see Section D3)
Содержание FLR 1000
Страница 1: ...FLR 1000 1000BR 1000ST Series Flow Sensors Meters For Liquids...
Страница 2: ......
Страница 25: ...pg 25 of 33 FLR 1000 BR FLR 1000 ST For Liquids 1 4 Compression Fittings Shown...
Страница 26: ...pg 26 of 33 FLR 1000 ST I For Liquids 1 4 Fittings Shown...
Страница 27: ...pg 27 of 33 FLR 1000 D For Liquids 1 4 Acetal Fittings Shown...
Страница 28: ...pg 28 of 33 FLR 1000 BR D FLR 1000 ST D For Liquids 1 4 Compression Fittings Shown...
Страница 32: ...pg 32 of 33...
Страница 33: ...M4574 0807 pg 33 of 33...