8-46
STRAIN
0
8
Scaling to
Display
Engineering
Units
again press the ‘MAX’ button to make your selection. After making
all of your choices, in submenu 1, enter these selections by pressing
the ‘MENU’ button which will momentarily flash “STORED”, on
the display. If the item selected in submenu 1 has a submenu 2
associated with it, press the ‘MIN’ button to enter the submenu 2.
Select your choice by pressing the ‘MAX’ button and enter your
choice by pressing the ‘MENU’ button.
Pressing the ‘RESET’ button one time brings you back one menu item,
pressing the ‘RESET’ button two times returns you to the run mode
WARNING! this also results in a hard reset of your peak and valley
and setpoints.
Proceed through the flow chart below entering the values listed in
submenu 1 and 2 columns.
MENU
SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF
Skip past
L2CNF
Skip past
L3CNF
Skip past
L4CNF
Skip past
INPUT
BRIDGE
Meter’s input
RDG.CNF RDG.1=0
2-point scaling IN.SC.OF
RDG.2=0
Active decimal point
RDG.3=0
Normal display brightness
RDG.4=1
Leading zeros suppressed.
RDG.5=
Disregard not used
RDG.6=
Disregard not used
RDG.7=0
External hard reset
vs only peak reset
RDG SC
1.00000
Preset for IN.SC.OF
RDG OF
0000000
Preset for IN.SC.OF
continued next page