9-25
Meter Function Menus
9
9.2.17 OT.SC.OF (Output Scale and Offset)
Output scale and offset is used to calibrate your optional analog output to
correspond to the engineering units you desire.
MENU
MIN/MAX/
BUTTON
MIN/MAX *
MENU **
MAIN
BUTTON
BUTTON
MENU
SUB MENU 1
SUB MENU 2
DESCRIPTION
“OT.SC.OF”:
OUTPUT SCALE & OFFSET
READ 1
Item #1 of Coordinate #1.
000000.
Enter the first value displayed
(“000000.”)
by the meter.
OUTPT1
Item #2 of Coordinate #1.
00.0000
Enter first desired
(“04.0000”)
output value.
READ 2
Item #1 of Coordinate #2.
000000.
Enter the second value
(“100000.”)
displayed by the meter.
OUTPT2
Item #2 of Coordinate #2.
00.0000
Enter second
(“20.0000”)
desired output value.
* The
‘MIN’
button allows you to sequence through “READ 1”, “OUTPT1”,
“READ 2”, and “OUTPT2” headings.
The
‘MAX’
button sends you to the value corresponding to “READ 1”,
“OUTPT1”, “READ 2”, and “OUTPT2” so you can change it (go to the SUB
MENU 2 item).
** The
‘MIN’
button allows you to step through the digits of the applicable
number being changed.
The
‘MAX’
button changes the value of the digit to be displayed.
The
‘MENU’
button stores the selected values for each input required in
“OT.SC.OF”. After the last value (“READ 2”) has been entered and the
‘MENU’
button is pressed, the meter display will advance to “BAUD”.
Every underlined item is the factory preset value.