8-33
8
Scaling to
Display
Engineering
Units
STRAIN
0
submenu 2 associated with it, press the ‘MIN’ button to enter the
submenu 2. Select your choice by pressing the ‘MAX’ button and
enter your choice by pressing the ‘MENU’ button.
Pressing the ‘RESET’ button one time brings you back one
‘MENU’ item, press the ‘RESET’ button two times to return to the
run mode.
WARNING! this also results in a hard reset of your peak and valley
and setpoints.
Proceed through the following flow chart entering the values listed
in submenu 1 and 2 columns.
MENU
SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF
Skip past
L2CNF
Skip past
L3CNF
Skip past
L4CNF
Skip past
INPUT
BRIDGE
Meter’s input
RDG.CNF RDG.1=0
2-point scaling IN.SC.OF
RDG.2=0
Active decimal point
RDG.3=0
Normal display brightness
RDG.4=1
Leading zeros suppressed
RDG.5=
Disregard not used
RDG.6=
Disregard not used
RDG.7=0
External hard reset vs
only peak reset
RDG SC
1.00000
Preset for IN.SC.OF
RDG OF
0000000
Preset for IN.SC.OF
IN CNF
INP.1=0
60 Hz ac power
INP.2=0
Slow reading
(S2A jumper omitted)
continued next page