12-5
Analog Output Options
12
12.6 FILTER CONFIGURATION “FIL.CNF” VALUE TO BE TRANSMITTED
ON ANALOG OUTPUT
1.
Press the
‘MENU’
button until the display shows “FIL.CNF”.
2.
Press the
‘MAX’
button until the display shows “FIL.3=0”, “FIL.3=1”,
“FIL.3=2”, or “FIL.3=3”.
3.
“FIL.3=0” Transmits the unfiltered value of your signal input.
“FIL.3=1” Transmits the filtered value of your signal input.
“FIL.3=2” Transmits the recorded PEAK (“HI RDG”) value(s).
“FIL.3=3” Transmits the recorded VALLEY (“LO RDG”) value(s).
4. Press the
‘MENU’
button to store your selection or changes and the
meter will advance to “FIL TI” (Filter Time Constant). Press the
‘RESET’
button two times to return to the run mode.
12.7 WIRING/CONNECTIONS
WARNING:
Do not connect ac power to your meter until you have
completed all input and output connections. Failure to do so may result in
injury! This device must only be installed electrically by specially trained
electrician with corresponding qualifications.
Figure 12-2. Isolated Analog Output Board Wiring Connections.
P5
1
2
3
Analog Voltage
0-10Vdc
Analog Current
4-20mA
Analog Return
+
+