7.4
Digital input functions
On the DP400S, digital inputs can be enabled by configuring the parameter 95
Digital input 1
and the parameter 100
Digital Input 2.
•
Enable outputs
: Enable operation of relays and analog outputs
•
Hold
: Lock the conversion.
•
Tare zero (AI)
: Set to zero the process value (tare function).
•
Alarm reset:
if one or more alarms are selected with manual reset and alarm
conditions are no longer present, closing the digital input it is possible to
restore the output related to alarm.
•
Totalizer reset
: if totalizer function is active, using the digital input it is
possible to reset the counter.
•
Peaks reset
: Reset min. peak/max. peak/peak-to-peak values
Sum total
: if sum function is active, using the digital input it is possible to
increase the Sum counter adding the process value.
•
Sum reset
: if the sum function is active, using the digital input it is possible to
reset the Sum counter.
•
Config. lock
: if the digital input is active it is not possible to enter configuration
or to modify the setpoints.
•
Gross/net TARE
: switches from the gross to the net weight visualization and
vice versa.
Selecting
Digital input 1
or
Digital input 2
on the alarm parameters, the
related relays will activate together with the digital input; functions selected on
parameters 95 and 100 will continue to work.
7.5
Peak values
Maximum
25.6 kg
Minimum
15.7 kg
Peak-Peak
9.9 kg
Esc
Rst
The DP400S is provided with a page for the
visualization of peak values: max. peak, min.
peak and peak-to-peak of analog input.
Keeping pressed “Rst” it is possible to reset
the visualized values.