
User manual - DP400S -
15
Key
Short pressure
Long pressure (1 s)
“>0<”
This function is enabled only if the
parameter 135 ->0<- is selected as Enabled.
Executes the reset of visualized weight.
This reset is kept also at switching off if the
parameter 13 Store is selected as Enabled.
“TARE”
This function is enabled only if the
parameter 135 TARE is selected as Enabled.
Executes the switch from gross to net
weight and vice-versa.
Ex.: once put the container on the scale,
display will show its weight. Press TARE to
visualize the net weight at 0 (on the upper
left side of display appears NET). Putting
the weighing item into the container,
display will show the product weight.
This operation can be repeated several
consecutive times.
Net/gross value is lost at switching off.
“Esc”
Back to process
visualization
“
”
Switch to menu
visualization
7.3
Calibration function
This function allows to activate a field calibration of linear input for strain-gauge
or potentiometer. Different calibration options are available and may be
selected by parameter, according to type of sensor/transducer and its relevant
application. To activate calibration procedure, enter configuration mode, select
the parameters group “Analog Inp.”, select Calibration parameter and chose
one of the available options pressing “Sel”. Following calibration options are
available:
Содержание DP400TP
Страница 53: ...User manual DP400S 53 ...
Страница 54: ...54 DP400S User manual ...
Страница 55: ...User manual DP400S 55 ...