SAFETY CONSIDERATIONS

This device is marked with the international Caution symbol.   It is important to read 
this manual before installing or commissioning this device as it contains important 
information relating to Safety and EMC (Electromagnetic Compatibility).

Unpacking & Inspection

Unpack the instrument and inspect for obvious shipping damage. Do not attempt to 
operate the unit if damage is found.

 This instrument is a panel mount device protected in accordance with Class I of 

 

EN 61010 (115/230 AC power connections). Installation of this instrument should be done 
by Qualified personnel. In order to ensure safe operation, the following instructions should 
be followed.

This instrument has no power-on switch. An external switch or circuit-breaker shall be 
included in the building installation as a disconnecting device.  It shall be marked to 
indicate this function, and it shall be in close proximity to the equipment within easy reach 
of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor 
(Earth wire), and it shall meet the relevant requirements of IEC  947–1 and IEC 947-3 
(International Electrotechnical Commission). The switch shall not be incorporated in the 
mains supply cord.

Furthermore, to provide protection against excessive energy being drawn from the mains 
supply in case of a fault in the equipment, an overcurrent protection device shall be 
installed.

  The Protective Conductor must be connected for safety reasons. Check that 

the power cable has the proper Earth wire, and it is properly connected. It is not 
safe to operate this unit without the Protective Conductor Terminal connected.

  Do not exceed voltage rating on the label located on the top of the 

instrument housing.

  Always disconnect power before changing signal and power connections.

  Do not use this instrument on a work bench without its case for safety 

reasons.

  Do not operate this instrument in flammable or explosive atmospheres.

  Do not expose this instrument to rain or moisture.

  Unit mounting should allow for adequate ventilation to ensure instrument 

does not exceed operating temperature rating.

  Use electrical wires with adequate size to handle mechanical strain and 

power requirements. Install without exposing bare wire outside the connector 
to minimize electrical shock hazards.

EMC Considerations

  Whenever EMC is an issue, always use shielded cables.

  Never run signal and power wires in the same conduit.

  Use signal wire connections with twisted-pair cables.

  Install Ferrite Bead(s) on signal wires close to the instrument if 

EMC problems persist.

Содержание DP301AN

Страница 1: ...DP301AN DC VOLTMETER DP301AN AC VOLTMETER e mail info omega com For latest product manuals www omegamanual info Shop online at omega com User sGuide...

Страница 2: ...rves the right to alter specifications without notice Servicing North America U S A Headquarters Omega Engineering Inc Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Cana...

Страница 3: ...r 3 2 9 Environmental 3 2 10 Mechanical 3 3 0 MECHANICAL ASSEMBLY AND INSTALLATION 4 3 1 Panel Mounting 4 4 0 POWER AND SIGNAL 6 4 1 Changing Operating Voltage 6 4 2 Power and Signal Connections 6 4 3...

Страница 4: ...sions 4 Figure 3 2 Exploded View 5 Figure 4 1 Main Board 6 Figure 4 2 Rear Terminal 6 Figure 5 1 Display Board Jumper Locations 10 Figure 5 2 Main Board Jumper Locations 11 Figure 5 3 Solder Switch Lo...

Страница 5: ...incorporated in the mains supply cord Furthermore to provide protection against excessive energy being drawn from the mains supply in case of a fault in the equipment an overcurrent protection device...

Страница 6: ...rement with readout from 0 to 1 999 1 2 MODEL DP301AN AC Model DP301AN AC for AC power line monitoring has a screw terminal barrier strip for signal and power This model uses a half wave average conve...

Страница 7: ...z CMV SIG LO to SIG GND 1 0 Vp CMR AC GND to SIG GND 120 dB DC to 60 Hz CMV AC GND to SIG GND 1500 Vp per HV test 354 Vp per IEC spacing 2 4 ANALOG TO DIGITAL CONVERSION Technique Dual slope average v...

Страница 8: ...56 V dc isolated to 300 Vp Power consumption 2 4 W nominal 2 9 ENVIRONMENTAL Operating temperature 0 to 60 C Storage temperature 40 to 85 C Relative humidity 95 at 40 C non condensing 2 10 MECHANICAL...

Страница 9: ...nough to allow for the panel thickness 3 Slide the two slide retainers towards the rear of the unit and remove them 4 Observe the case dimensions previously shown From the front of the panel insert th...

Страница 10: ...5 3 1 PANEL MOUNTING continued WARNING High voltage is present when using model AN AC with the 650 V AC option Figure 3 2 Exploded View...

Страница 11: ...he voltmeters are factory set for 115 V ac 15 operation Power and signal can be connected using the D1 or D4 option Figure 4 2 Rear Terminal OPERATING VOLTAGE JUMPER INSTALLATION 230 V ac 15 Option C1...

Страница 12: ...High 6 High 2 N Low 5 Low 3 Ground 4 Signal GRND DC D1 2 DC 17 High C DC return 16 Low D4 1 No connection 6 High 2 DC 5 Low 3 DC return 4 Signal GRND Connector Power Signal D4 1 L High 6 High 2 N Low...

Страница 13: ...N J1 Connector D1 Option A 1 N C B 2 AC LO DC PWR C 3 N C D 4 N C E 5 N C F 6 DIG GND H 7 199 9 DP J 8 19 99 DP K 9 1 999 DP L 10 LAMP TEST M 11 4 7 V N 12 EXC P 13 EXC R 14 HOLD S 15 Spare T 16 SIG L...

Страница 14: ...4 4 DIGITAL TEST SIGNALS Display HOLD instructs the meter to maintain a display of the last reading To activate HOLD connect J1 pin 14 R to J1 pin 11 M LAMP TEST instructs the meter to light the minu...

Страница 15: ...en close solder switches as indicated Remove all push on jumpers not used 5 1 DECIMAL POINT SELECTION Figure 5 1 Display Board Jumper Locations Decimal Point S1 Alternate Decimal Point Configuration U...

Страница 16: ...ure 5 3 on the following page for solder switch locations Input Configuration Solder Switches Push on Jumpers Remove Jumper Open Close S2 S3 Wire 200 mV dc A A 2 V dc B 20 V dc B C 200 V dc B D 200 V...

Страница 17: ...12 5 3 BLANKING THE LEAST SIGNIFICANT DIGIT To blank the least significant digit open solder switch A on the circuit side of the display board Figure 5 3 Solder Switch Locations...

Страница 18: ...age equal to 95 of the high end of the range selected 4 Locate the Span potentiometer R23 in Figure 3 2 Adjust until the display reads 1900 1 count 6 2 AC VOLTAGE RANGES 1 Follow steps 1 and 2 from ab...

Страница 19: ...14 7 0 DRAWINGS Figure 7 1 Assembly Diagram...

Страница 20: ...15 NOTES...

Страница 21: ...ty or otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipmen...

Страница 22: ...Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Equipm...

Отзывы: