1.0 

GENERAL INFORMATION

The two versions of voltmeters are different pinout versions of the same 3 1/2 
digit panel meter for applications which require a compact, quality dc voltmeter.  
Standard features include HOLD function, selectable decimal point location and 
LAMP TEST.

Each model contains the circuitry for all ranges.  The range is changed by 
opening and closing the appropriate switches on the printed circuit board via 
installation of push-on jumpers.

1.1 

MODEL DP301AN

Model DP301AN is pin compatible with newer voltmeters DP302, DP3001 and 
DP3002 series. 
Models DP301AN has a resolution of 1 part in ±1999 counts, or 0.05% of full 
scale. By using an external dc voltage reference, any of these ranges can be 
configured for 3 wire ratio measurement with readout from 0 to 1.999.

1.2 

MODEL DP301AN-AC

Model DP301AN-AC, for AC power line monitoring, has a screw terminal barrier 
strip for signal and power. This model uses a half wave average converter 
calibrated to read RMS. It offers input ranges of 199.9 and 650 V ac, with 
resolution of 100 mV and 1 V, respectively. Maximum rated input in either range is 
650 V RMS or 920 Vp.

2.0 

SPECIFICATIONS

2.1 

ANALOG INPUT 

Model "AN" 
Signal ranges  

 

±199.9mV     ±1.999V     ±19.99V     ±199.9V

Bias current   

 

50 pA           50 pA  

 5 pA 

      1 pA

Maximum input  

 

130 Vp         130 Vp 

 250 Vp       250 Vp

Resolution   

 

100 uV         1 mV 

 10 mV        100 mV

Input resistance  

 

1 G

 

       1 G

 

 1.1 M

       1.0 M^N8,50

Configuration 

   Differential, bipolar

Zero 

    Automatic

Span adjustment     ±5% 
Model "AN-AC" 
Signal ranges  

 

199.9 V ac 

650 V ac

Bias current             1 pA          

1 pA

Overvoltage protection  

650 V ac 

650 V ac

Resolution           100 mV        

1 V ac

Input resistance        

1 M

   

1 M

  

Configuration           

Half wave average, RMS calibrated

Zero     

   Automatic

Frequency               

47 to 1000 Hz

1

Содержание DP301AN

Страница 1: ...DP301AN DC VOLTMETER DP301AN AC VOLTMETER e mail info omega com For latest product manuals www omegamanual info Shop online at omega com User sGuide...

Страница 2: ...rves the right to alter specifications without notice Servicing North America U S A Headquarters Omega Engineering Inc Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Cana...

Страница 3: ...r 3 2 9 Environmental 3 2 10 Mechanical 3 3 0 MECHANICAL ASSEMBLY AND INSTALLATION 4 3 1 Panel Mounting 4 4 0 POWER AND SIGNAL 6 4 1 Changing Operating Voltage 6 4 2 Power and Signal Connections 6 4 3...

Страница 4: ...sions 4 Figure 3 2 Exploded View 5 Figure 4 1 Main Board 6 Figure 4 2 Rear Terminal 6 Figure 5 1 Display Board Jumper Locations 10 Figure 5 2 Main Board Jumper Locations 11 Figure 5 3 Solder Switch Lo...

Страница 5: ...incorporated in the mains supply cord Furthermore to provide protection against excessive energy being drawn from the mains supply in case of a fault in the equipment an overcurrent protection device...

Страница 6: ...rement with readout from 0 to 1 999 1 2 MODEL DP301AN AC Model DP301AN AC for AC power line monitoring has a screw terminal barrier strip for signal and power This model uses a half wave average conve...

Страница 7: ...z CMV SIG LO to SIG GND 1 0 Vp CMR AC GND to SIG GND 120 dB DC to 60 Hz CMV AC GND to SIG GND 1500 Vp per HV test 354 Vp per IEC spacing 2 4 ANALOG TO DIGITAL CONVERSION Technique Dual slope average v...

Страница 8: ...56 V dc isolated to 300 Vp Power consumption 2 4 W nominal 2 9 ENVIRONMENTAL Operating temperature 0 to 60 C Storage temperature 40 to 85 C Relative humidity 95 at 40 C non condensing 2 10 MECHANICAL...

Страница 9: ...nough to allow for the panel thickness 3 Slide the two slide retainers towards the rear of the unit and remove them 4 Observe the case dimensions previously shown From the front of the panel insert th...

Страница 10: ...5 3 1 PANEL MOUNTING continued WARNING High voltage is present when using model AN AC with the 650 V AC option Figure 3 2 Exploded View...

Страница 11: ...he voltmeters are factory set for 115 V ac 15 operation Power and signal can be connected using the D1 or D4 option Figure 4 2 Rear Terminal OPERATING VOLTAGE JUMPER INSTALLATION 230 V ac 15 Option C1...

Страница 12: ...High 6 High 2 N Low 5 Low 3 Ground 4 Signal GRND DC D1 2 DC 17 High C DC return 16 Low D4 1 No connection 6 High 2 DC 5 Low 3 DC return 4 Signal GRND Connector Power Signal D4 1 L High 6 High 2 N Low...

Страница 13: ...N J1 Connector D1 Option A 1 N C B 2 AC LO DC PWR C 3 N C D 4 N C E 5 N C F 6 DIG GND H 7 199 9 DP J 8 19 99 DP K 9 1 999 DP L 10 LAMP TEST M 11 4 7 V N 12 EXC P 13 EXC R 14 HOLD S 15 Spare T 16 SIG L...

Страница 14: ...4 4 DIGITAL TEST SIGNALS Display HOLD instructs the meter to maintain a display of the last reading To activate HOLD connect J1 pin 14 R to J1 pin 11 M LAMP TEST instructs the meter to light the minu...

Страница 15: ...en close solder switches as indicated Remove all push on jumpers not used 5 1 DECIMAL POINT SELECTION Figure 5 1 Display Board Jumper Locations Decimal Point S1 Alternate Decimal Point Configuration U...

Страница 16: ...ure 5 3 on the following page for solder switch locations Input Configuration Solder Switches Push on Jumpers Remove Jumper Open Close S2 S3 Wire 200 mV dc A A 2 V dc B 20 V dc B C 200 V dc B D 200 V...

Страница 17: ...12 5 3 BLANKING THE LEAST SIGNIFICANT DIGIT To blank the least significant digit open solder switch A on the circuit side of the display board Figure 5 3 Solder Switch Locations...

Страница 18: ...age equal to 95 of the high end of the range selected 4 Locate the Span potentiometer R23 in Figure 3 2 Adjust until the display reads 1900 1 count 6 2 AC VOLTAGE RANGES 1 Follow steps 1 and 2 from ab...

Страница 19: ...14 7 0 DRAWINGS Figure 7 1 Assembly Diagram...

Страница 20: ...15 NOTES...

Страница 21: ...ty or otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipmen...

Страница 22: ...Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Equipm...

Отзывы: