OMEGA Model CYD201/CYD208 User’s Manual
5-6
Service
5.7.2
A/D Converter Calibration
NOTE: Current source calibration
must be performed before A/D converter
calibration.
1. Allow 30 minute warm-up to achieve rated specifications.
2. Perform current source calibration.
3. Configure 125 k
resistor as shown in Figure 5-4. Set Model 208s to
channel 1.
4. Wait 10 seconds for resistor voltage to settle.
5. Press CAL ENABLE, then within 2 seconds press UNITS. The display
window shows "-CAL-" for approximately 15 seconds.
6. Repeat procedure with 25 k
resistor.
7. For Model CYD208, repeat procedure for other 7 channels. See
Figure 2-2 for pin assignments.
5.8
SERIAL INTERFACE CABLE AND ADAPTERS
12
3
4
5
6
1
2
3
4
5
6
BLACK
RED
GREEN
YELLOW
RxD
Gnd
Gnd
TxD
Figure 5-5. Model CYD200-J10 RJ-11 Cable Assembly Wiring Details
13
12
11
10
9
8
7
6
5
4
3
2
1
25
24
23
22
21
20
19
18
17
16
15
14
= NOT
USED
6
5
4
3
2
1
DB-25 CONNECTOR
RJ-11
RECEPTACLE
RxD
Gnd
TxD
Figure 5-6. Model CYD200-D RJ-11 to DB-25 Adapter Wiring Details
Содержание CYD201
Страница 1: ...M789 038A 15 June 1999...
Страница 6: ...OMEGA Model CYD201 CYD208 User s Manual iv Table of Contents This Page Intentionally Left Blank...
Страница 22: ...OMEGA Model CYD201 CYD208 User s Manual 2 10 Installation This Page Intentionally Left Blank...
Страница 28: ...OMEGA Model CYD201 CYD208 User s Manual 3 6 Operation This Page Intentionally Left Blank...
Страница 36: ...OMEGA Model CYD201 CYD208 User s Manual 4 8 Remote Operation This Page Intentionally Left Blank...
Страница 44: ...OMEGA Model CYD201 CYD208 User s Manual 5 8 Service This Page Intentionally Left Blank...
Страница 48: ...OMEGA Model CYD201 CYD208 User s Manual 6 4 Options and Accessories This Page Intentionally Left Blank...
Страница 57: ...NOTES...
Страница 58: ...NOTES...
Страница 60: ......