OMEGA Model CYD201/CYD208 User’s Manual
Installation
2-9
2.8
INITIAL POWER UP SEQUENCE
The test sequence below occurs at power up.
1. All display segments light.
2. The unit displays “-201-” or “-208-”.
3. The instrument begins normal operation. Units currently selected flash.
Model CYD208s also indicate the current channel selected.
2.9
POWER UP ERRORS
On power up, the CYD201/CYD208 checks internal memory. If a problem
exists, an error message displays on the front panel of the instrument.
“Er01” indicates a hardware problem in the instrument memory. This error is
not user-correctable. First perform the procedure in Paragraph 5.3. If
unsuccessful, then call the factory.
“Er02” indicates a soft error in the instrument memory. To correct this error,
close dip switch 1 on the rear panel for at least 5 seconds, then open it.
Follow the calibration procedure described in Paragraph 5.3 after an error 2
reset.
“OL” indicates a voltage input overload. This can be caused by an open
sensor or diode sensor wired backwards.
Содержание CYD201
Страница 1: ...M789 038A 15 June 1999...
Страница 6: ...OMEGA Model CYD201 CYD208 User s Manual iv Table of Contents This Page Intentionally Left Blank...
Страница 22: ...OMEGA Model CYD201 CYD208 User s Manual 2 10 Installation This Page Intentionally Left Blank...
Страница 28: ...OMEGA Model CYD201 CYD208 User s Manual 3 6 Operation This Page Intentionally Left Blank...
Страница 36: ...OMEGA Model CYD201 CYD208 User s Manual 4 8 Remote Operation This Page Intentionally Left Blank...
Страница 44: ...OMEGA Model CYD201 CYD208 User s Manual 5 8 Service This Page Intentionally Left Blank...
Страница 48: ...OMEGA Model CYD201 CYD208 User s Manual 6 4 Options and Accessories This Page Intentionally Left Blank...
Страница 57: ...NOTES...
Страница 58: ...NOTES...
Страница 60: ......