OMEGA Model CYD201/CYD208 User’s Manual
Installation
2-7
2.6
SENSOR CURVE DEFINITION
To display accurate temperature, select a response curve that matches the
installed sensor. There are seven standard curves stored within the Model
CYD201/CYD208 numbered 0 through 6 (see Appendix A). Different curves
may be assigned to each channel of the Model CYD208. Find the unit
factory curve configuration inside the front cover of this manual. Curve 6
(CY-7 Curve 10) is the standard curve configuration unless specified
differently upon order.
To determine current curve selection, press and hold UNITS and turn on (1)
the rear panel SET switch (DIP switch 3). Release UNITS key. The CYD201
displays the curve number in the display window. The Model CYD208
displays the curve number in the display window and the channel number in
the channel window. To display curves for other channel numbers in the
Model CYD208, press CHANNEL to scroll through the eight channels.
To change the curve, press UNITS. The instrument scrolls through curves 0
through 6. In the Model CYD208, press CHANNEL to select other channels,
then press UNITS to scroll through the curves.
After the new curve selection, turn the SET switch on the rear panel off (0).
The unit returns to normal operation.
Table 2-2. Model CYD201/CYD208 Temperature Curves
Curve No.
Range (K)
Description
0
1
2
3
4
5
6
0 – 324.9
0 – 324.9
0 – 324.9
0 – 324.9
0 – 324.9
0 – 324.9
0 – 474.9
DT-500DI-8B
DT-500DI-8A
DT-500DRC-D
DT-500TDC-E1
CTI Curve C
DT-500DI-8C
CY-7 Curve 10
2.7
RACK MOUNTING
The Models CYD201 and CYD208 can install in a standard “size” 1/4 panel
EIA rack space. If you ordered a CYD208-DIN rack mounting adapter, follow
the installation instructions below. See Figure 2-3.
1. Remove front feet on bottom of unit and attach lower rack piece by
threading two of the four screws provided into the front feet holes.
2. Locate the two mounting hole access covers on the top of the unit.
Attach the other rack with the remaining screws.
Содержание CYD201
Страница 1: ...M789 038A 15 June 1999...
Страница 6: ...OMEGA Model CYD201 CYD208 User s Manual iv Table of Contents This Page Intentionally Left Blank...
Страница 22: ...OMEGA Model CYD201 CYD208 User s Manual 2 10 Installation This Page Intentionally Left Blank...
Страница 28: ...OMEGA Model CYD201 CYD208 User s Manual 3 6 Operation This Page Intentionally Left Blank...
Страница 36: ...OMEGA Model CYD201 CYD208 User s Manual 4 8 Remote Operation This Page Intentionally Left Blank...
Страница 44: ...OMEGA Model CYD201 CYD208 User s Manual 5 8 Service This Page Intentionally Left Blank...
Страница 48: ...OMEGA Model CYD201 CYD208 User s Manual 6 4 Options and Accessories This Page Intentionally Left Blank...
Страница 57: ...NOTES...
Страница 58: ...NOTES...
Страница 60: ......