OMEGA Model CYD201/CYD208 User’s Manual
Service
5-5
5.7
RECALIBRATION
OMEGA calibrates and certifies thermometers to original factory
specifications for a reasonable fee. You can recalibrate the thermometer to
original specifications, but OMEGA will not warrant these calibrations.
Recalibration requires a digital voltmeter (DVM) with 4½ digit resolution or
better; and 25 k
and 125 k
precision resistors with ±0.01% tolerance or
better.
Recalibration involves current source and analog/digital (A/D) converter
calibration. A/D calibration erases all SoftCal™ compensations; perform it
after current source calibration, not before.
5.7.1
Current Source Calibration
1. Allow 30 minute warm-up to achieve rated specifications.
2. Configure 125 k
resistor as shown in Figure 5-4. Set Model CYD208 to
channel 1.
3. Connect DVM voltage leads across the resistor and adjust the I ADJ
trimpot (located on rear panel) until DVM displays a voltage of 1.2500
volts ±100 microvolts.
4. Remove DVM and resistor.
Model CYD201
Model CYD208
NOTE: Do not use a voltmeter for A/D Converter Calibration.
Figure 5-4. Calibration Connections
Содержание CYD201
Страница 1: ...M789 038A 15 June 1999...
Страница 6: ...OMEGA Model CYD201 CYD208 User s Manual iv Table of Contents This Page Intentionally Left Blank...
Страница 22: ...OMEGA Model CYD201 CYD208 User s Manual 2 10 Installation This Page Intentionally Left Blank...
Страница 28: ...OMEGA Model CYD201 CYD208 User s Manual 3 6 Operation This Page Intentionally Left Blank...
Страница 36: ...OMEGA Model CYD201 CYD208 User s Manual 4 8 Remote Operation This Page Intentionally Left Blank...
Страница 44: ...OMEGA Model CYD201 CYD208 User s Manual 5 8 Service This Page Intentionally Left Blank...
Страница 48: ...OMEGA Model CYD201 CYD208 User s Manual 6 4 Options and Accessories This Page Intentionally Left Blank...
Страница 57: ...NOTES...
Страница 58: ...NOTES...
Страница 60: ......