- 4 - 

2

 

Installation   

2.1  Installation of the binocular viewing head 

1)  Loosen the head thumb lock screw on the top of the microscope body and remove 

the plastic cover on the top. 

2)  Remove the cap on the dovetail of the binocular viewing head. 
3)  Seat the dovetail completely of the viewing head into the socket on the top of the 

microscope body and tighten the head thumb lock screw. 

Caution

:   

Do not release the viewing head from your hand grip until you are sure the viewing 
head is installed securely.   

2.2  Installation of the eyepieces   

1)  Remove the protective caps from the eyepiece tubes.   
2)  Insert the eyepieces into the eyepiece tubes.     

2.3  Installation of the objectives 

1)  Adjust the coarse focus knob until the mechanical stage is at its lowest position. 
2)  Turn the caps counter-clockwise to remove them from the nosepiece. 
3)  Take the objectives out from the plastic cases and turn each one clock-wise into the 

holes on the nosepiece. Install the 4X objective into the nosepiece first. Then in a 
counter-clockwise direction, rotate the nosepiece and install each succeeding higher 
magnification objective as shown in 

Fig. 1

   

 
 
 
 
 
 
 
 
 

Note: 

  Inspect the objectives frequently for dirt or oil; clean if necessary. 

  Use the 10X objective to initially focus the image of your specimen.   

  When changing the objective magnification, rotate the objective nosepiece until 

you hear a “click” sound or have a clear "in position" feeling. This ensures the 
objective is centered in the optical light path. 

2.4  Installation of the glass filter 

1)  Swing out the color filter holder under the condenser. 
2)  Place the filter into the holder as shown in 

Fig. 2

, swing the holder in.   

   

 
 

Fig.1 

Содержание M827S Series

Страница 1: ...User Manual Binocular Biological Compound Microscope Model M827S Series MicroscopeNet com...

Страница 2: ...1 Component Illustration 3 2 Installation 4 3 Operation 7 4 Specifications 10 5 Optional Parts 11 6 Troubleshooting Guide 13 7 Darkfield Condenser Installation and Operation Instructions 15 8 Phase C...

Страница 3: ...icroscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto the stage objective or any other component disconnect the power cord immediately and wi...

Страница 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Страница 5: ...Screw 19 Condenser Lock Thumb Knob 4 Nosepiece 12 Microscope Body 20 Condenser Focus Knob 5 Objective 13 Focus Tension Ring 21 Focus Stop Lever 6 Slide Holder 14 Coarse Focus Knob 22 Abbe Condenser 7...

Страница 6: ...mechanical stage is at its lowest position 2 Turn the caps counter clockwise to remove them from the nosepiece 3 Take the objectives out from the plastic cases and turn each one clock wise into the h...

Страница 7: ...t You may also loosen the two screws on the cover to adjust the position of the bulb to get centered and even brightness See Fig 3 7 Screw the cover on Caution Before you turn over the microscope be s...

Страница 8: ...no power is available 2 8 Installation of the camera optional may not included in your package 1 Take off the eyepiece from the eyepiece tube 2 Insert the camera into the eyepiece tube and then connec...

Страница 9: ...pupillary distance While observing with both eyes hold the left and right eyepiece tubes then slide the tubes in and out The interpupillary distance is correct when the left and right fields of view c...

Страница 10: ...ctive nosepiece to seat the 100X objective to the observing position until you hear a click sound 4 After observing the specimen use the lens cleaning paper to clean the 100X objective lens gently and...

Страница 11: ...ocus tension has been pre set at the factory If the mechanical stage drops by itself rotate the tension adjustment ring Fig 11 situated between the coarse focus knob and microscope body on the power s...

Страница 12: ...adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides w focus stop Minimum fine focusing adjustment at 0 002mm range 28mm Mechanical Stage Double layer dimension 5 1 2 x 5 1...

Страница 13: ...00X 1 25 oil 0 5 160 0 17 spring 37mm A191BOIL Oil 1 36 1 25 Plan 100X 1 25 oil 0 5 160 0 17 spring w iris diaphragm 37mm 2 Cameras Model Resolution Operating System Software A1502 640 x 480 0 3MP MS...

Страница 14: ...ring Achromatic 100X with built in phase plate spring oil five positions 10 for 10X phase contrast objective 20 for 20X phase contrast objective 40 for 40X phase contrast objective 100 for 100X phase...

Страница 15: ...too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Diopter adjustment is not set proper...

Страница 16: ...brightness The condenser is not in the right position Lower or raise the condenser slightly to the position that the specimen is the brightest A191BOIL A191BD only Make sure the oil contact the bottom...

Страница 17: ...ondenser up or down slightly to get the best darkfield viewing Note The dry darkfield condenser is used with the dry objectives only The dry darkfield condenser works with the 4X 10X 40X objectives Th...

Страница 18: ...jective you will need to drop the immersion oil and adjust the iris ring Fig 16 to get proper brightness and contrast of view field 8 After observing the specimen with 100X objective use the lens clea...

Страница 19: ...ighten the thumb screw g Insert the light collector back onto the base and turn until it s tightened h Re adjust the condenser height to its normal position i Re adjust the mechanical stage height to...

Страница 20: ...e condenser control knob to raise the condenser to the highest position e Loosen the condenser lock thumb screw Fig 20 pull down the condenser and take it off f Insert the condenser annular ring disk...

Страница 21: ...onto the condenser as shown in Fig 22 c Loosen the condenser lock thumb screw as shown in Fig 23 take off the original condenser from the holder d Insert the phase contrast condenser into the condense...

Страница 22: ...g is in focus and the dark ring is visible i If the two ring images are not coincided as shown in Fig 24 c adjust the two centering screws on the condenser ring plate till the two rings are coincided...

Отзывы: