Appendix
75
HIGH FLOW INSUFFLATION UNIT UHI-3
VOLUME
indicator
Display
method
• 0.0 – 19.9 L, 0.1 L step
• 20 – 999 L, 1 L step
Reset
Pressing the reset button will return the display to
0.0 L.
Smoke evacuation function
(optional)
• When the foot switch (optional) is depressed,
insufflation and suction are interlocked to
maintain abdominal pressure at the set pressure
while smoke evacuation is performed.
• Inoperative when “LOW” is selected as
insufflation mode.
• Inoperative when the abdominal pressure is
below 3 mmHg.
Classification
(electro-medic
al equipment)
Type of
protection
against
electric shock
Class
I
(3-pin power cord)
Degree of
protection
against
electrical
shock of
applied part
Type CF applied part:
Insufflation tube (MAJ-590, MH-397)
Suction tube (MAJ-591)
Weight
10 kg
Dimensions
295 mm (W) × 340 mm (D) × 150 mm (H) (housing
dimensions)
Power
requirements
Voltage (AC)
100 – 240 V
Frequency
50/60 Hz
Input
150 VA
Voltage
fluctuation
Within
±
10%
Suction pump
Adjust to
−
400 to
−
300 mmHg.
Year of
manufacture
The year of manufacture is the second digit of the
serial number.
Item
Specification
1212345