32
Chapter 3 Installation and Connection
HIGH FLOW INSUFFLATION UNIT UHI-3
4.
Connect the sterilized suction tube to the connector on the trocar that is
positioned close to where the smoke is being created.
5.
Insert the suction tube’s small diameter tube all the way into the groove of
the CO
2
suction control pinch valve on the front panel of the UHI-3 (see
Figure 3.10). To prevent the tube from collapsing, there must be at least
10 cm between the differential diameter connector and the pinch valve
groove (see Figure 3.11).
Figure 3.10
Figure 3.11
6.
Connect the suction tube via a suction container to the operating room’s
suction pump.
To suction installation
CO
2
SUCTION
Trocar, etc.
Differential diameter connector
10 cm