24
Chapter 3 Installation and Connection
HIGH FLOW INSUFFLATION UNIT UHI-3
3.2
Connecting to an AC mains supply
• Connect the UHI-3 main unit directly to a hospital grade AC
outlet (wall mains outlet).
• Securely plug in the power cord so it will not be accidentally
dislodged during the operation.
• Never apply excessive force to the power cord, such as
bending, straining, twisting or squeezing.
• Do not use a 3-pin plug to 2-pin plug adapter as this will not
properly ground the UHI-3.
• Always use the power cord provided with the UHI-3.
1.
Confirm that the power is OFF.
2.
Connect the power cord plug directly to a 3-pin hospital grade AC outlet
(wall mains outlet) which meets the power requirements indicated on the
electrical rating plate on the rear of the UHI-3 (see Figure 3.1).
Figure 3.1
Properly
grounded wall
mains outlet
Power plug